52
Continued
Main
Code
Sub
Code
ERROR/ALARM indicator
Description
Error remedy method
Refer to section
Lighting
interval
Switching off
condition
91
5C5
0.25 seconds
Power
Communication between
Disc Drive LSI 1b
(IC400: BD-48 board)
and BRIDGE (IC700
internal: BD-48 board)
cannot be established
during startup.
Turn off and on the unit, and then perform again firmware update.
•
The unit power may have been shut down during firmware update.
If the same error occurs, a drive fault is likely.
5C7
0.25 seconds
Power
Failure in Disc Drive LSI
1b (IC400: BD-48 board)
firmware upgrade was
detected during startup.
66F
0.25 seconds
Power
A firmware error of
SYCPU (IC700 internal:
BD-48 board) was
detected.
Acquire the drive/driver log, and then turn off and on the unit.
•
In the case no error occurs:
The problem is fixed. Use the unit as it is.
•
If the same error occurs repeatedly:
Send the acquired drive/driver log to your local Sony Sales Office/Service
Center.
-
695
0.25 seconds
Power
Communication between
SYCPU (IC700 internal:
BD-48 board) and Disc
Drive LSI 0a (IC100:
BD-48 board) cannot be
established during startup.
Or, failure in Disc Drive
LSI 0a (IC100: BD-48
board) firmware upgrade
was detected during
startup.
Turn off and on the unit, and then perform again firmware update.
•
The unit power may have been shut down during firmware update.
If the same error occurs, a drive fault is likely.
Содержание PDW-U4
Страница 7: ...4 Frame Wiring 151 Revision History 152 ...
Страница 24: ...21 Hold the shaded portions Keep away a screwdriver Objective lens Actuator ...
Страница 147: ...144 Frame Assembly 2 310 309 310 309 311 311 311 311 307 306 305 302 308 304 311 306 311 303 301 311 307 ...
Страница 155: ...152 Revision History Date History Contents 2020 10 1st Edition 9 932 764 01 ...
Страница 156: ...PDW U4 SY PDW U4 CN J E 9 932 764 01 Sony Corporation Printed in Japan 2020 10 08 2020 ...