1-20
•
Do not operate or subject the PVA-1500xx to temperatures beyond the
published operating and storage temperature specifications.
•
Wear electrical safety gloves.
•
Wear eye protection.
•
Wear fall protection where required.
•
Assume that metal surfaces are energized unless proven otherwise.
•
Isolate the PV source circuit under test from the inverter, and from other PV
source circuits, before making any connections to the test device (PV module,
string, or array).
•
Always pause the measurement sequence using the LED-illuminated pushbutton
switch on the I-V Unit before connecting or disconnecting the test leads of the
PVA-1500xx.
•
Do not use the PVA-1500xx to test devices that produce more than the
instrument’s specified maximum current and voltage.
•
Connect the test leads to the test device (PV module, string, or array) with the
correct polarity.
•
Protect the primary test lead connectors of the PVA-1500xx by installing
connector-saver jumpers. Replace the connector-saver jumpers when they have
reached 100 connections.
•
Special 1500v test leads and alligator clips are provided with the PVA-1500xx.
Do not substitute other test leads. Also, do not use the test leads from a PVA-
1000S, which are rated only to 1000v.
WARNING
Do not remove instrument covers. There are no user serviceable parts within.
Operation of the instrument in a manner not specified by Solmetric may result in
personal injury or loss of life.
•
Do not use the I-V Unit if it is damaged. Always inspect for damage before
using.
•
Keep PV connectors away from soil and dirt. Always inspect PV connectors for
damage and dirt before connecting. Do not use if damaged.
•
Do not use the I-V Unit if it is performing abnormally. Contact Solmetric for
guidance or return the I-V Unit to the factory for service.
Содержание PVA-1000S
Страница 12: ...xii This page is intentionally left blank ...
Страница 68: ...1 Getting Started 2 56 Over Voltage Warnings ...
Страница 116: ......
Страница 128: ......
Страница 133: ...7 Interpreting Measured I V Curves 7 5 5 4 Rounder knee 6 ...
Страница 148: ......
Страница 152: ...8 Translation of I V Data to Standard Test Conditions 8 4 This page is intentionally left blank ...