4.5.3 Reference positions adjust
Position the teacher to the right rear corner of the patient rack area.
4.5.3.1 Adjustment of the reference left position
Click the
icon in the
[Needles Adjust]
dialog to open the
[Ref.Left Adjust]
dailog.
Note: It
’
s not necessary to adjust the such position for MAGLUMI 1000.
Fig. 4.5.3.1-1: [Ref.Left Adjust] dialog
To set the stepsize-bar can adjust value of the stepsize. Set bigger
stepsize firstly , then press the
icon to move the pipettor needle
downward .When the needle get close to target posision, set stepsize
smaller. Adjust the needle in X-axis and Y-axis direction With the
、
、
、
icon , making the needle tip just above
the teacher’s center point .Then press
、
icon to make the
needle tip higher than the teacher’s center point by 0.5mm. Save the
parameters with
icon
,
then press
<OK>
to
exit.
Requirements
:
1. Needle tip just above the teacher’s center point
2. Needle tip higher than the teacher’s center point by 0.5mm.