Safety Notes
7.2 Electric Shock caused by Improper Grounding
Use Power Cable furnished with the MAGLUMI
®
Analyzer. Use of
Power Cable other than the specified one can cause Electric Shock.
8 Deterioration of users state of health
8.1 Injury to Hand caused by Needle or Mechanism
While running the Analyzer, do not insert your hand or anything into
the Apparatus, your hand, finger or arm can be injured
Capped sample tubes can cause a crash of the needle , therefore
make sure to uncap all sample tubes before the sample racks are
introduced into the MAGLUMI
®
Analyzer. If a needle crash occurs,
the
procedure defined in the Chapter “Sample Station” is to be followed.
The MAGLUMI
®
Analyzer may never been run without the mounted
Safety Shield in front of the pipetting area.
Cleaning and disinfecting actions can cause injuries by the needles,
therefore cleaning and disinfecting work is to be done only if the
MAGLUMI
®
Analyzer is not into service.
The Cleaning and Disinfection Procedure in this Operating
Instructions are strictly to be followed.
Fig. 8.1-1: Warning on Bar over Pipetting Area
Page 6
Operating instructions