Chapter 9: Troubleshooting
Home Use Operation Manual
9-5
Other
PRObleM
POSSIble CAUSe
CORReCTIVe ACTIOn
Cannot change alarm limits
after pushing the menu
key.
Monitor is in Home “
K
” mode.
This is proper operation.
No sounds from monitor.
Audio paused (silenced).
Defective Speaker.
Push silence key.
Do not use to monitor
patients. Contact your
authorized service
representative.
Clock shows erroneous
values.
Clock battery disconnected or
discharged.
Contact your authorized
representative.
Display shows
sr
ERR
No sensor attached to the
oximeter.
Defective sensor or patient cable.
Attach a sensor to the
oximeter.
Try a new sensor. If the
problem persists, contact
your authorized service
representative.
Display shows
24
ERR
EEPROM checksum Error
Contact your authorized
service representative.
Display shows
25
ERR
Trend Checksum Error
Contact your authorized
service representative.
Display shows
ERR
Internal oximetry failure.
Contact your authorized
service representative.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......