Chapter 7: Optional Docking Station and Printer
7-8
Home Use Operation Manual
Mode
3
- Graphic Trend Mode prints in a chart format. See figure 7-9.
TREND DATA
ID: 009______________________
Trend Time Scale=20 Sec/Point
0 ...........
PI
...........20
0
► ► ► ► ►
Pulse Rate(bpm)
► ► ► ► ►
254
50
100
▪ ▪ ▪ ▪ ▪ ▪ ▪
SpO2(%)▪ ▪ ▪ ▪ ▪ ▪ ▪
04
00:08
00
Figure 7-9: Sample Graphic Trend
nOTe: While printing trends, the printer will pause for approximately 20 seconds after
every 10 seconds of printing. This allows the thermal print head to cool. A small gap
may appear on each printout when printing resumes.
nOTe: If the printer runs out of paper during a printout, the printer will not automatically
continue printing when a new roll of paper is installed. After installing a new roll of
paper, the printout must be restarted.
nOTe: Invalid SpO
2
data and SpO
2
values between 0 and 50 will be printed as 50.
Trend Data Condition Flags
Each time a trend interval occurs and data is stored, certain conditions, active at that time, are
stored with the data. These are indicated by symbols on the print out.
COnDITIOn
SyMbOl
DeSCRIPTIOn
Parameter Alarm
Indicates a parameter has met or exceeded its alarm limit.
Artifact
Artifact indicator informs user of excess motion, noise or
other signal information that the algorithm interprets as
potentially non-physiological.
This flag is an indicator that pulse rate data has changed and
now may be invalid.
Small Pulse
Indicates the signal strength is ≤ 3.
Check Sensor
Indicates a problem with sensor placement or that no sensor
is plugged into the sensor connector. No valid parameter
data is available.
Searching too Long
Only displayed if the oximeter has not detected a valid
pulse at any time after power up. Indicates that the oximeter
has searched for more than 20 seconds but no pulse was
detected.
Lost Pulse
Indicates the oximeter has searched for more than 20
seconds, a finger is detected in the sensor, but a previously
detected pulse is no longer present.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......