Chapter 5: Operating Instructions
Home Use Operation Manual
5-3
Installing the batteries
The oximeter uses four (4) standard “AA” alkaline, IEC Type LR6, cells (Figure 5-2A) or a custom
rechargeable Lithium-Ion (Li+) battery pack (WW1090 - Figure 5-2B).
WARnInG! Inspect battery terminals for corrosion or contamination. The monitor may
not operate properly or could fail to alarm if battery terminals are corroded or
contaminated. Do not use until battery terminals have been properly cleaned
and repaired.
WARnInG! Check the expiration date for the batteries. The monitor may not operate
properly or could fail to alarm if expired batteries are used. Do not use until
proper batteries can be obtained.
WARnInG! Remove the batteries prior to long term storage.
WARnInG! Do not allow the patient to handle the device if the battery door has been
removed, except while installing new batteries.
To install/replace the batteries:
Depress the battery door tab near the center of the oximeter and slide the cover off toward
1.
the bottom of the monitor.
2a. If using “AA” (LR6) alkaline batteries: Install the negative end of each battery first,
compressing the battery terminal spring until the positive terminal clears the positive tab.
Press the battery down into place.
nOTe: Dispose of spent batteries in compliance with your institution’s guidelines and local
ordinances.
2b. If using WW1090 Li-Ion (Li+) rechargeable battery pack: Align the battery pack so that the
metal connectors line up with the connectors in the WW1030 battery compartment. Push
the WW1090 Lithium-Ion (Li+) rechargeable battery pack straight in to place.
Replace the battery door by sliding the cover back until the latch clicks.
3.
Figure 5-2A: AA (LR6) Alkaline Battery Placement
+
+
+
+
Batteries
Battery
Door
Figure 5-2B: Rechargeable Lithium-Ion Battery Pack
Rechargeable Battery Pack
Battery
Door
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......