Chapter 5: Operating Instructions
Home Use Operation Manual
5-7
Turning On the Monitor
Figure 5-6: Turning On the Monitor
To turn on the monitor, press the
x
key. Before using the monitor, check the following at power
up:
All LEDs light.
•
The monitor beeps briefly.
•
The monitor’s software revisions (main, battery PIC, oximeter) are momentarily displayed.
•
Displays “
•
--HI
” if SpO
2
high alarm limit is OFF, and not in Sleep Mode.
Displays “
•
AL OFF
” (Alarms Off) if in Sleep Mode.
Displays “
•
HS YES
” if High Sensitivity Mode is ON.
Displays Operation Mode indicator:
•
Patient Record Number for Clinician Mode.
•
“
•
SLP
” for Sleep Mode.
“
•
H
” for Home Mode.
After a few seconds the % SpO
2
value, pulse rate, pulse signal strength and PI bar graphs should
be shown. If not, see
Chapter 9: Troubleshooting
for help.
WARnInG! Verify that all leDs (light emitting diodes) on the display light up upon startup
of the device.
WARnInG! Verify that the monitor sounds a short tone upon startup of the device. If no
tone is heard, the speaker may be damaged. Do not use to monitor patients
until the monitor has been repaired.
If the WW1030 detects an error during power up, “
ERR
” will display in the pulse rate section of
the display. A numeric error code will display in the SpO
2
section. See
Chapter 9: Troubleshooting
for more information.
Press this key to turn
the monitor on and off.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......