Chapter 6: Alarms
Home Use Operation Manual
6-5
low Priority Alarms
A low priority alarm warns of an abnormal monitor condition.
WARnInG! During a low priority alarm the monitor cannot measure the patient’s SpO
2
or pulse rate. you must immediately check the patient’s condition. After the
patient’s condition has been checked, you must correct the low priority alarm.
See
Correcting the Alert
in
Chapter 9: Troubleshooting
for help.
A low priority alarm is activated when:
The sensor is not connected to the monitor.
•
The monitor searches too long for an initial reading.
•
The sensor is not properly attached to the patient.
•
The sensor is connected to the monitor, but no finger is detected in the sensor.
•
A fault is detected in the sensor or sensor cable.
•
During a low priority alarm:
The low priority alarm tone sounds if not silenced. The low priority alarm sound is a single
•
tone pair of beeps with a 20 second pause.
The Alarm Indicator lights yellow.
•
Each time new trend data is saved. It is marked with an alert active flag.
•
The alarm condition is sent to the data output port.
•
Alarms active at the time of a real time data output are indicated on the optional WW1026
•
thermal printer.
Alarm Summary
TyPe
DISPlAy InDICATOR
AlARM
InDICATOR
eFFeCTS
AUDIO
High Priority
Alarms
Numbers corresponding
to violated alarm will
flash twice per second.
Flashes red
twice per
second.
Overrides any
other alarm and
pulse beeps.
2 bursts of 5 tones
each, repeated
every 7 seconds.
Medium
Priority Alarms
(low battery,
internal comm
error)
For low battery: battery
charge indicator flashes
one yellow segment
For internal
communication error:
display shows error code.
Flashes
yellow once
every 2
seconds.
Overrides a Low
Priority Alarm
and pulse
beeps.
Set of 3 tones
each, repeated
every 20 seconds.
Low Priority
Alarms
Dashes display in the
SpO
2
and pulse rate
areas.
Lights
yellow
Overrides pulse
beeps.
Single tone pair of
beeps, repeated
every 20 seconds.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......