TPN

Total Parenteral Nutrition Delivery Mode

Model 6100 and 6101

Ambulatory

Infusion

Pumps

P A T I E N T   I N F O R M A T I O N

Содержание CADD Prizm VIP 6100

Страница 1: ...TPN Total Parenteral Nutrition Delivery Mode Model 6100 and 6101 Ambulatory Infusion Pumps PAT I E N T I N F O R M AT I O N ...

Страница 2: ...r AC Adapter 11 The Main Screen 13 Description of the Infusion Profile 14 How to Use the HELP Key Å 15 Starting the Pump 16 Stopping and Restarting the Pump 17 Removing a Cassette 18 Attaching a Cassette 19 Priming the Tubing and Starting the Pump 22 Inserting the Tubing into the Air Detector 24 What if I drop or hit the pump 25 Alarms and Messages 26 ...

Страница 3: ... parenteral nutrition solution or other fluids at a constant rate Delivery may taper up or gradually increase at the beginning of delivery Delivery may also taper down or gradually decrease at the end of delivery The pump stores programmed information and historical information that your clinician needs for your specific therapy Your clinician will instruct you on the proper use of this pump This ...

Страница 4: ...under delivery will increase and could on occasion be significant You must regularly compare the volume remaining in the syringe to the pump s displayed values such as Reservoir Volume or Given to determine if under delivery is occur ring and if necessary contact your clinician Do not use rechargeable NiCad or nickel metal hydride NiMH batteries Do not use carbon zinc heavy duty batteries They do ...

Страница 5: ...art with 21 73xx You must use a CADD Extension Set with Anti Siphon Valve or a CADD Administration Set with either an integral or add on Anti Siphon Valve to protect against delivery inaccura cies and unregulated gravity infusion that can result from an improperly attached cassette For detailed instructions and warning pertaining to Medi cation Cassette Reservoirs or CADD Administration Sets pleas...

Страница 6: ...rther instructions Cautions Do not operate the pump at temperatures below 2 C 36 F or above 40 C 104 F Do not store the pump at temperatures below 20 C 4 F or above 60 C 140 F Do not store the pump with a Medication Cassette Reservoir or CADD Administration Set attached Do not expose the pump to humidity levels below 10 or above 90 relative humidity Do not store the pump for prolonged periods with...

Страница 7: ...ck accessory jack or Air Detector Port area Do not clean the pump with acetone other plastic sol vents or abrasive cleaners Do not use the pump in the presence of flammable anes thetics or explosive gasses Use only Smiths Medical MD accessories as using other brands may adversely affect the operation of the pump ...

Страница 8: ...voir or CADD Administration Set that attaches to the pump Green Light Blinks every 3 seconds when the pump is running and deliver ing fluid Amber Light Flashing pump is stopped or an alarm exists Steady pump is inoperable call clinician Display Shows information and messages After a short time the display turns itself off to save power Press any key to turn the display back on Battery Compartment ...

Страница 9: ...ey explains what you see on the display Used by the clinician Advances you from one screen to the next and silences some alarms Í This key is not used in the TPN Delivery mode Used by the clinician Lets you answer yes to a question on the pump s display Î Lets you answer no to a question on the pump s display ...

Страница 10: ...not use rechargeable NiCad or nickel metal hydride NiMH batteries Do not use carbon zinc heavy duty batteries They do not provide sufficient power for the pump to operate properly which could result in death or serious injury Always have new batteries available for replacement If power is lost non delivery of drug will occur and depend ing on the drug being administered could result in death or se...

Страница 11: ...ttery door and slide the battery door forward Remove the used battery 4 Match the and markings on the new battery with the mark ings on the pump Insert the battery The pump will beep if the battery is inserted correctly 5 Replace the battery door The pump will power up auto matically 6 Start the pump by pressing 7 When you see Start the pump press NOTE If you put the battery in backwards the displ...

Страница 12: ...go through an electronic self test and the pump will beep at the end of the power up sequence All of the display indicators the software revision level and each parameter will appear briefly CAUTION Do not store the pump for prolonged periods of time with the battery installed Battery leakage could dam age the pump ...

Страница 13: ... as a backup Otherwise you won t be able to start the pump Your Power Pack or AC Adapter may or may not have the connector shown 1 Open the Power cover on the side of your pump 2 Line up the red mark on the Power Pack cord connector with the red mark on the pump s Power jack 3 Push the connector forward until it stops Do not twist or turn the connector Pull lightly on the cord directly behind the ...

Страница 14: ...r to the Power Pack input jack Or if you want to operate the pump on AC power you can attach the AC Adapter directly to the pump s Power jack Detaching the Power Pack 1 Grasp the connector 2 Pull the connector back using a straight steady motion DO NOT twist or turn the connector 2000 03 21 D Zurn EPS Compnts NoCallouts w Cone ...

Страница 15: ... key lets you look at the pump s program Status of Infusion Period See the next page If the pump is stopped this will show STOPPED Status of Reservoir Volume The Main Screen The following screen is what you will see on the pump s display most of the time It is called the Main Screen and shows the following ...

Страница 16: ...may not have programmed a taper up or a taper down If there is no taper up delivery will start at the plateau rate If there is no taper down you will still hear 9 beeps at the end of the infusion profile but the rate will immediately drop from the plateau rate to the KVO rate Then when you stop the pump it will automatically reset to the beginning of the infusion profile RUNNING Taper up RUNNING T...

Страница 17: ...ppear The symbol in the lower right corner means there are more help screens Press Å again to see the next help screen To page through all the help screens press Å repeatedly This will take you through all the help screens then back to the original screen you had questions about To return to the original screen you had questions about without going through all the help screens press This is the ca...

Страница 18: ...ism An undetected upstream occlusion may result in under or non delivery of drug and depending on the drug being adminis tered could result in death or serious injury Air embolism can result in death or serious injury 1 Press 2 When Start the pump appears press Starting pump will appear The pump will automatically review and display the prepro grammed settings RUNNING will appear on the main scree...

Страница 19: ...VO delivery the Infusion Profile and the Reservoir Volume will automatically be reset This means that when you restart the pump it will start from the beginning of the Infusion Profile 1 Press 2 When Stop the Pump appears press You will see two screens if you stopped the pump during KVO delivery STOPPED will appear on the main screen and the amber light will blink Stop the Pump Press Y or N TPN 62...

Страница 20: ...connect the tubing from your access device as instructed by your clinician 3 Use a coin to unlatch the cas sette Insert the coin into the slot and turn clockwise until the latch ing button pops out NOTE If you cannot turn the latching button the cassette may be locked Contact your clinician 4 Remove the cassette from the pump 5 Discard the Medication Cassette Reservoir or CADD Administration Set a...

Страница 21: ...erious injury For detailed instructions and warnings pertaining to the Medication Cassette Reservoir or CADD Administration Set please refer to the instructions for use supplied with those products Frozen medication must be thawed at room temperature only Do not heat the Medication Cassette Reservoir in a microwave oven as this may damage the medication the Medication Cassette Reservoir or cause l...

Страница 22: ...th the solid dot WARNING Attach the cassette the part of the Medication Cassette Reservoir or CADD Administration Set that at taches to the pump properly An improperly attached or detached cassette could result in unregulated gravity infu sion of medication from the fluid container or a reflux of blood which could result in death or serious injury 5 Gently twist and pull on the cas sette to make s...

Страница 23: ...d Press 7 Reset Reservoir Volume to may appear Press to reset the Reservoir Volume to the originally programmed value This will also reset the Infusion Profile Reset Reservoir Volume to 3000 0 ml Press Y or N High Volume Admin set latched NEXT to continue Infusion Profile has been reset ...

Страница 24: ...atheter This could result in over delivery of medication or air embolism which could result in death or serious injury 3 Press and hold the key until the tubing is fully primed or until priming stops 4 This screen will appear if you release the key or if priming automatically stops If the tubing is not yet fully primed press the key and repeat step 3 If you are finished prim ing press the Î key Pr...

Страница 25: ...Open all clamps 6 Press to start the pump Starting pump will appear The pump will automatically review and display the programmed settings RUNNING will appear on the main screen the green light will blink and fluid delivery will begin as programmed by your clinician WARNING Ensure that the entire fluid path is free of all air bubbles before connecting to your catheter to prevent air embolism Air e...

Страница 26: ...inician has instructed you to use the Air Detector open the Air Detector door and thread the tubing through the groove 2 Close the door making sure the tubing does not get pinched or kinked 3 Follow your clinician s instructions for connecting the tubing to your access site Open all clamps WARNING Ensure that the entire fluid path is free of all air bubbles before connecting to your catheter to pr...

Страница 27: ...on the pump Gently twist and pull on the cas sette to make sure it is still firmly attached Check the battery door to make sure it is still firmly attached If the cassette or battery door are loose or damaged do not use the pump Immediately stop the pump close the tubing clamp and contact your clinician WARNING If the pump is dropped or hit inspect the pump for damage Do not use a pump that is dam...

Страница 28: ...s table or press Å for help When you see Take this action The 9 volt battery is low but the pump is operable Change the 9 volt battery soon Install a new 9 volt battery The pump will not run with a de pleted 9 volt battery A good battery must be installed even when an external source of power is connected A two tone alarm is sounding and the amber light stays on The 9 volt battery is depleted Inst...

Страница 29: ... fluid container to the pump Check for a kink a closed clamp or an air bubble in the tubing between the fluid container and pump Press to stop the pump and silence the alarm for 2 minutes then remove the obstruction and press to restart the pump There may be a kink in the tubing or a clamp may be closed Unkink the tubing or open the clamp and the pump will resume delivery You may press to stop the...

Страница 30: ... as appropriate The Reservoir Volume has reached 0 0 ml Press to stop the alarm Then change the fluid container if appropriate The Reservoir Volume value is low indicating that the level of fluid in the fluid container is low Fluid is not flowing from the fluid container to the pump Check for a kink a closed clamp or an air bubble in the tubing between the fluid container and pump Press to stop th...

Страница 31: ...s not flowing from the fluid container to the pump Check for a kink a closed clamp or an air bubble in the tubing between the fluid container and pump Press to stop the pump and silence the alarm then remove the obstruction and press to restart the pump There is a problem with the pump Close the tubing clamp and remove the pump from service Contact your clinician ...

Страница 32: ...ctions Pump placement during bathing showering ____________ ______________________________________________________ Pump placement during sleep _________________________ ______________________________________________________ Storage of medication ________________________________ ______________________________________________________ ...

Страница 33: ...31 Notes ...

Страница 34: ......

Страница 35: ......

Страница 36: ...mes trademarks or service marks of their respective owners The products described are covered by one or more of the following U S Patent Nos 4 559 038 4 565 542 4 650 469 5 181 910 5 338 157 5 364 242 5 485 408 5 531 697 5 531 698 5 538 399 5 540 561 5 564 915 5 567 136 5 567 119 5 695 473 Model 6101 only other patent s pending Smiths Medical MD Inc St Paul MN 55112 U S A 1 800 426 2448 www smiths...