To Clean SPIDER2
1. Remove all single use components and accessories. Discard single use components.
Refer to cleaning and sterilization instructions for accessories: 79-99-04 Cleaning and
Sterilization Instructions.
2. Wipe entire SPIDER2 with mild detergent / disinfectant, taking care to avoid fluid entry
into Switch Drape or Foot Pedal Receptacles.
3. Wipe battery with mild detergent / disinfectant.
4. Dry SPIDER2 with a cloth or allow to air dry.
5. DO NOT immerse any portion of SPIDER2 in liquid. If moisture enters the equipment it
can cause corrosion, and damage electrical and/or mechanical components.
6. DO NOT autoclave SPIDER2.
7. DO NOT use ethylene oxide to sterilize SPIDER2.
Storage
1. Turn OFF and DEPRESSURIZE SPIDER2 by disconnecting the Foot Pedal and/or Switch
Drape and pressing the Power Switch to the DEPRESSURIZE Position for 1 second (Figure
2). The Power Switch will then return to the OFF/STORAGE Position.
2. Always depressurize and then turn off the SPIDER2 when not in use. Failure to do so will
reduce the life expectancy of the device.
Maintenance
Only trained individuals should perform maintenance on SPIDER2. Contact Smith &
Nephew to determine which maintenance can be performed on site.
SPIDER2 Specifications*
*Specifications are approximate and may vary from unit to unit or as a result of power
supply fluctuations
Model
REF 72203299 SPIDER2
Size (fully extended)
Length: 1495 mm Width: 180 mm
Height: 200 mm
Weight
30 lbs [13.6 kg] (with battery installed)
Approval
CSA International
CAN/CSA-C22.2 No.601.1-M90
UL 60601-1
IEC 60601-1
Equipment type
Type B Applied Part
Power Supply
Internally Powered 24V 15A
Battery Type
NiCd 24V 1.8Ah
Battery Model
REF 72203301 SPIDER2 Battery Pack
Enclosure Protection
IPX1 Drip-proof equipment
Environmental Conditions
Operation Storage & Transportation
Temperature
10°C - 40°C -40°C - 70°C
Relative Humidity
30% - 75% 10% - 100%
Atmospheric Pressure
70kPa - 106kPa 50kPa - 106kPa
Disposal
EU: Not for general waste
Contact Smith & Nephew for appropriate methods of
disposal.
SPIDER2 Battery Charger Specification
Approval
Intertek
CAN/CSA Std C22.2 No. 60950-1
UL 60950-1
IEC 60950-1
Power supply
100-240 VAC, 50-60 Hz, Max 0.9 A
Ordering Information
SPIDER2 Accessories*
REF
Description
72203301
SPIDER2 Battery Pack
7210570
Piggy Back
7210749
90° Connector
7210577
Traction Accessory
7210571
Shoulder Connection Bar
7210576
Lateral Shoulder Connection Bar
7210572
Elbow Connection Bar
72200660
Wrist Arthroscopy Accessory
72202932
Ankle Arthroscopy Accessory
72203235
SPIDER Leg Accessory (Left)
72203236
SPIDER Leg Accessory (Right)
72203239
3D SPIDER Connector
* SPIDER2 is packaged with:
(1) SPIDER2
(1) SPIDER2 Battery Pack
(1) SPIDER2 Battery Charger
(3) Power Cords (Type A - NEMA 1-15, Type C - CEE 7/16 Europlug, Type G - BS 1363).
Contact Smith & Nephew for additional power cords, if required.
Stabilization Kits
72203833
SPIDER2 Shoulder Stabilization Kit
7210573
Shoulder Stabilization Kit
7210574
Elbow Stabilization Kit
72201550
Small Wrist Stabilization Kit
72201551
Medium Wrist Stabilization Kit
72201552
Large Wrist Stabilization Kit
72202931
Ankle Stabilization Kit
72203238
Leg Stabilization Kit
72203300
Switch Drape
Warranty
This product has a standard 12-month parts and labor warranty against manufacturing
defects.