3.4 FLUSHING THE UNIT WITH XYLENE
If needed, (before a longer period of non-use of the instrument, a transport and re-
installation on a new site and similar situations) you should start a system flush with
xylene. (At initial setup after receiving a new MCS I, after re-installation on a new site
and transports it is also strictly necessary to ensure, no remaining mounting media can
get hardened out and stick inside the mounting media system!)
From the operation screen you can press „Setup“, and then „Arm&Disp“ and then
„Dispenser“ to get into the needed pump menu.
Be sure, you emptied and replaced the purge cap before operating.
Put a container / glass with xylene into the mounting media bottle position, insert the
tube, put an empty purge cap (for mounting media waste) and start the system flush
with the key “Cleaning“. The unit now starts pumping xylene from your xylene
container to fill and clean the complete system. You will have to exchange / empty the
purge cap after the MCS I asks you to check xylene and Pertex® 3 times during this
cycle. You should restart the “Cleaning“ run perhaps for a second and third time after it
has finished. In cases of system cleaning after longer non-use of the MCS I, be sure to
have 3-4 runs before changing to Pertex® again and then start 2 runs with Pertex®
(Point 3.4). (To ensure a clean and primed unit, you should have a system flush in
intervals of every 2
nd
or 3
rd
mounting media exchange).
After a system flush the MCS I is prepared for filling up with Pertex®! (step 3.5)