16
Manual_MCS_I_2021-10-002.00_EN.docx
3.7 SETUP DATA
3.7.1 MOUNTING MEDIA VOLUME, COVER GLASS SIZE, ETC
To set the position of the start and end points of the mounting media line and the
volume of the media, Slee specially focused on simplicity for users. The volume of
Pertex® can be changed in 5µl steps anytime during the process using the two upper
keys on the right side of the display between 30 and 150µl. (In operation screen) For
more elementary settings, you may go to the „Setup“ menu by pushing the „Setup”
key, where all possibilities for an individual setting are open.
By entering „Arm&Disp“ you go on with “Dispenser” into the setting functions for the
„mounting media“. Here, you can choose between „Position“ and the cleaning or
bubble free functions.
By choosing „Position“ you will get the possibility, to set the start and end positions of
the mounting media – according to your used cover glass and type of samples. You can
change the distance from the left slide edge to the start point of the line (X1) with the
left two upper keys at the display and the length of the mounting media line (X2) with
the right upper keys at the display in 1mm steps.
Because the media spreads after cover glass is applied, it normally is sufficient, to enter a
value for X1 from left slide edge to cover glass edge and add about +5mm and for X2
of about cover glass length minus 10mm. Otherwise too much media might be pressed
out to the sides.
The “delay” time should only be changed, if you have problems with the start point of
the line.(~150ms).
For first tests Slee recommends to set the volume to the lowest value (30µl) and to
increase this volume during the test runs till results are optimized. This can be done
easily during the process with „” and „volume-„ in the operation screen. You
don’t need to change to the „Setup“ menu!
Get out of any screen – back to the last one – easily by pressing exit in any screen.
All changes will be directly stored inside the program.