Manual_MCS_I_2021-10-002.00_EN.docx
17
3.7.2 SETUP “ARM” (SPEED AND PRESSURE)
To set the pressure and / or the speed of the coverglass head while bringing up the
coverglass onto the slide, go into “Setup” / “Arm & Disp” / “Arm” and adjust the
pressure and speed parameters from values between “10” and “200” to find the best
solution between reducing air-bubbles and being fast.
Best is, to have the MCS I adjusted by your local Slee sales or service representative. Slee
strictly recommends “automat star” coverglass from Knittel with a thickness of 0,12 to
0,19mm to receive best results without having problems of sticking to each other, that
can lead to crashes and damage of slides in bad cases, as the MCS I is working very fast
and has to bring all slides back into the basket, which may be impossible if too many
covers or splinters are placed there.
!!! Risk of broken slides and splinters of coverglass if ignored!!!
Be sure to guarantee a correct stock keeping of your coverglass! Avoid temperature
differences and high humidity! If still having such problems, there are some special
functions for Slee personal to activate a save mode of coverglass pickup. In these cases,
please call your local SLEE medical GmbH representative and ask for help in activating
and adjusting.
3.7.3 PROGRAMS
The MCS I can store up to 4 individual settings in different “programs”. Changes inside
the actual chosen program (indicated as “P#” in the lower left corner of the operation
screen – for example “P1”) will automatically be stored and used by the controller.
Mounting media length, volume, pressure or speed – all changes will only take effect in
the program, which you currently activated! This means, if you should need slower
speed and pressure for a special kind of samples, maybe due to air bubbles on the slides
for example, you may switch to another program no.# .For example “P2”. Now all
parameters changed to the values, which were actually stored in “P2”. You can now set
all parameters here to the same values, as in your “standard” program and only change
speed and pressure to a lower level. Finally set values are now automatically stored in
“P2” and it is easy now to switch the MCS I to these values, each time you run these
samples again, just by choosing “P2” again. Afterwards, you may switch back to your
standard / routine values by turning back to “P1”. The actual chosen program is
indicated in the main screen on the lower left side of the display (for example: “P1”). To
call up other programs, just turn to “SETUP” and choose another one by pressing
“P1(P2/P3/P4) change”. P1 changes from P1 to P2, then P3 or P4, then starts back at P1,
each time you hit this key one time.