Manual_MCS_I_2021-10-002.00_EN.docx
3.3 SWITCH ON / CHECK / INITIALIZATION
Connection and main switch “ON”
Connect the MCS I with your laboratory power supply.
Turn on the main switch (at the power connector on the
right side) and close both covers if open.
Press “ON/OFF” key at the display
With connected power cable the unit now is in standby
and can be switched to operation with the power key on
the display (left side, upper button).Press this power
switch and hold it for 2 seconds.
Eventually „power failure“
Quit the „power failure“ if shown. (This means, the unit
was not correctly shut down after last time running,
maybe due to a power blackout).
Check and confirm filling levels
Now check the levels of xylene cartridges and the
mounting media bottle. The machine has released the
hood locking now, so it's easy for you to open it and
have a closer look at cartridges and the bottle, and to
refill them, if necessary. Also check for the cover glass
waste container and the purge cap (mounting media
waste, which has to be replaced and emptied regularly).
The purge caps are available as consumables at Slee.
The MCS I asks you to confirm your check by pressing
“OK”.