6.2 TROUBLESHOOTING
Problem Possible
Reason
Values, such as pressure, speed or
mounting media volume suddenly
have changed.
→
someone changed values (by mistake?)
→
someone hit the „P# change“ key by mistake and changed to other
program no.# due to this
CG head cannot get up any cover
glass
→
cover glass not OK
→
vacuum pump defective
→
vacuum tube blocked (by mounting media or glass splinters
→
Height adjustment of CG head should be proceeded (SERVICE)
Dispenser doesn't bring up
mounting media (PERTEX) onto the
slides.
→
mounting media bottle run empty (no further media inside bottle /
tube not correct?)
→
Dispenser needle blocked (no xylene inside needle reservoir?)
→
the check valve inside system may be broken/blocked (SERVICE)
→
mounting media pump defective
→
pumping system not lubricated (xylene level for lubrication not OK?)
6.3 ERROR MESSAGES
The errors of the step motor axes and internal controller communication don't have a special screen with
user information. So errors with these units are shown as number codes on the screen. The numbers have
a special meaning. The first number shows the step motor-number, which created the error code. The
second number (mostly 1 or 2) shows more detailed information for Slee service people.
1.
stands for „reference error“ (sensor not reached due to blockage, broken motor, broken sensor,
cable or bad connection)
2.
stands for „runtime error“ (couldn't execute actual step)