Whatever CG-Head system you use, it is absolutely necessary to keep the vacuum pads
absolutely clean from any dirt or mounting media. For cleaning, you may pull the pads
off to clean them with xylene or suitable solvent or to replace them. Only put them back
when they are dry, otherwise the pads will stick to your coverglass when starting the
next process. Never clean the pads while still fixed to the CG-Head.
Correct placement of the pads to the vacuum tube is guaranteed by checking, if the
small nose on the pad is placed in the slot at the opposite. The pressure of the
separators of the CG-Heads are adjusted by a screw which can easily be turned more up
or down to find the optimal separation spring pressure. (See in the pictures on the
previous page).
In both versions, the pressure has to be exactly adjusted, according to the quality of the
used coverglass, humidity or thickness, to get the best results between best separation
and fewest broken glass problems.
The right picture shows the 1 vacuum pad
system which here is adjusted not too
hard. (separation pressure adjustment
screw adjusted more to the upper end)