Page 82
EN
PGA 3 Input of setup values, Table 1 from 2
Step
Key
Function
8
Select setup-value input menu
Press
function key F1
The setup-value input menu is jumped to
Use the
next step function key
to select the setup-value measuring point to be changed
9
Use the
numeric keypad
to enter the new setup value (pulses/minute)
If 0 pulses/minute are entered as the setup value for a measuring point, this measuring point is deactivated. If an erroneous
entry is made, the input can be rejected by pressing key S2.
Press the
ENTER
key
or
press key
S2
reject
10
Select data-save menu
Press
next step function key
until the data-save menu is jumped to
11
Store or reject changed setup values
The data-save menu calls on the operator to send the new setup values to the pulse meter for storage. When function key S1
is pressed the new setup values are transferred to the IPM12 pulse meter selected (in point 2) and stored.
Press
function key S1
(data storage)
or reject new setup values:
Press
function key S2
The main menu is jumped back to without the setup-value data being saved.
7. Operating
S1
Содержание IPM 12
Страница 47: ...Seite 47 DE...
Страница 51: ...Seite 51 DE 11 4 COMO Zentralschmiersystem berwachung 11 Anschlussbeispiele Men struktur...
Страница 53: ...Seite 53 DE...
Страница 101: ...Page 101 EN...
Страница 105: ...Page 105 EN 11 4 Overview COMO Lube System Monitoring 11 Connection example menu structure...
Страница 107: ...Page 107 EN...