Safety
2 Safety
2.1 Use
2.1.1 Intended use
The device has been designed and built exclusively for its intended
purpose as described in these instructions. The device is to be used
only for dynamic centric compression or tensile-loaded lift of a load.
It is intended for interior use only.
The device has been designed for the movement of medical
equipment, specifically medical procedure chairs and tables.
2.1.1.1 Product service life
The telescopic pillar CPMT is designed for a service life of 10 years in
a typical medical procedure equipment application.
Product service life of the telescopic pillar CPMT depends on the
stroke and load of the application. According to the L
10
lifetime of the
device, 56 000 cycles can be achieved in an average medical procedure
application (average load of 4000 N and average stroke of 150 mm)
Some reference values for different stroke and load conditions are
given in the table below.
Lifetime [cycles] reference at given average load and stroke
Average load F
m
4 000 N
5 000 N
6 000 N
CPMT1-X
S
0XXXFX-000
Stroke = 150 mm
42 000
23 000
14 000
Stroke = 300 mm
21 000
11 000
7 000
Stroke = 500 mm
12 000
7 000
4 000
CPMT1-X
M
0XXXFX-000
Stroke = 150 mm
56 000
30 000
18 000
Stroke = 300 mm
28 000
15 000
9 000
Stroke = 500 mm
17 000
9 000
5 000
6
www.lmotion.ru
Тел
.: (495)-921-34-60