Quality control
Motor quality control testing performed in a rewind shop or production facility could use
relatively higher voltages compared to maintenance testing of the same motor. In a quality
control testing situation, the test results are either pass or fail. Data is not trended and
diagnosis is not a goal. Rather, the insulation system of a newly-rewound motor must be
able to withstand test levels in accordance with IEEE and NEMA standards.
When conducting quality control tests, failures must be analyzed. For example, some
winding configurations will fail a winding resistance test with seemingly reasonable test
parameters even when the winding resistance is good. Knowledge of special windings is
important and can only be provided by those responsible for the winding.
Motor troubleshooting
In the case of a motor failing during service, the Baker AWA-IV can help determine root-
cause reasons for failure, and provide data that assists in making sound decisions about
refurbishment vs. replacement. Isolation of a phase-to-phase short, for example, can be
done with a surge test performed on the windings. A shorted winding will yield a waveform
substantially different from two good windings.
Motor shops can use the Baker AWA-IV to indicate where a problem has occurred,
pinpointing what needs to be repaired. Baker AWA-IV test results provide a tool for
recommendations to the customer.
Field coils
When testing field coils, follow the procedures outlined for tests upon single-phase motors,
two-terminal devices, and synchronous motors/generators. The recommended surge test
voltage for DC fields is 600 Volts.
If impedance of the coils is very low (fewer turns, generally form coils with very low
resistance), the surge tester standalone might not adequately test the coils. A bar-to-bar,
low-impedance test accessory from SKF may be needed.
Hi L in Baker AWA-IV 2 kV and Baker AWA-IV 4kV
Hi L is a technique that extends the range of the base surge test circuitry in 2 kV and 4 kV
models of the Baker AWA-IV. This circuitry, like all electric circuits, has design and
102
SKF Static Motor Analyzer—Baker AWA-IV User Manual
Special features of the Baker AWA-IV
Содержание AWAIV-12
Страница 1: ...SKF Static Motor Analyzer Baker AWA IV User manual ...
Страница 2: ......
Страница 14: ...xii SKF Static Motor Analyzer Baker AWA IV User Manual Table of Contents ...
Страница 16: ...2 SKF Static Motor Analyzer Baker AWA IV User Manual About this manual ...
Страница 28: ...14 SKF Static Motor Analyzer Baker AWA IV User Manual Baker AWA IV Instrument Overview ...
Страница 88: ...74 SKF Static Motor Analyzer Baker AWA IV User Manual Database management and maintenance ...
Страница 90: ...76 SKF Static Motor Analyzer Baker AWA IV User Manual Set up of the Baker AWA IV analyzer ...
Страница 91: ...SKF Static Motor Analyzer Baker AWA IV User Manual 77 Set up of the Baker AWA IV analyzer ...
Страница 92: ...78 SKF Static Motor Analyzer Baker AWA IV User Manual Set up of the Baker AWA IV analyzer ...
Страница 124: ...110 SKF Static Motor Analyzer Baker AWA IV User Manual Special features of the Baker AWA IV ...
Страница 166: ...152 SKF Static Motor Analyzer Baker AWA IV User Manual Typical winding faults ...
Страница 180: ...166 SKF Static Motor Analyzer Baker AWA IV User Manual ...
Страница 181: ...SKF Static Motor Analyzer Baker AWA IV User Manual 167 ...
Страница 182: ...168 SKF Static Motor Analyzer Baker AWA IV User Manual ...
Страница 194: ...180 SKF Static Motor Analyzer Baker AWA IV User Manual Using the Baker ZTX with Baker AWA IV analyzers ...
Страница 210: ...196 SKF Static Motor Analyzer Baker AWA IV User Manual ...
Страница 234: ...220 SKF Static Motor Analyzer Baker AWA IV User Manual ...
Страница 240: ...226 SKF Static Motor Analyzer Baker AWA IV User Manual Glossary ...
Страница 248: ...234 SKF Static Motor Analyzer Baker AWA IV User Manual Index ...
Страница 249: ......