Power
[W]
Modus
Duty
Cycle
[%]
Fre-
quenzy
[Hz]
Fiber
[µm]
Help Menu
Frenectomy
3,0
CW
320
Stretch the frenulum and stay in contact with
the fiber. Use brush stroke at the base to cut
through fibrous attachment.
Caution: For the tongue, protect the salivary
glands! Avoid bone contact during treatment!
Gingivectomy
3,0
CW
320
Gently shape the gingival tissue in contact with
the fiber.
Caution: Work in parallel to the tooth surface!
Gingivoplastic
3,0
PF
50
10
320
Gently shape the gingival tissue in contact with
the fiber.
Caution: Work parallel to the tooth surface!
Haemostasis
3,0
CW
320
Seal small blood vessel with gentle contact to
the tissue. Permeate larger vessels with fiber,
start laser and retract the fiber slowly. To co-
agulate larger vessel perform multiple treat-
ments.
Implant
Uncovery
3,0
CW
320
Stretch the tissue and use laser tip like an
scalpel to excise the tissue.
Caution: Avoid contact to implant and bone!
Incisions/
Excisions
3,0
CW
320
Stretch the tissue and use laser tip like an
scalpel to excise the tissue.
Operculec-
tomy
3,0
CW
320
Stretch the tissue and use laser tip like an
scalpel to excise the mucosa hood.
Gingival
Troughing
2,0
PF
50
20
320
Gently shape the gingival tissue in contact with
the fiber.
Caution: Work parallel to the tooth surface!
6.3
Further non-preset indications
Further non-preset applications which you may define individually and
medicate according to scientific publications in My Settings:
● Frenotomy
● Biopsy
● Laser assisted flap surgery
● Incisions and draining of abscesses
● Papillectomy
● Vestibuloplasty
● Excision of lesions
● Excision of hyperplasias
● Leukoplakia
● Crown lengthening
● Set hidden teeth free