8 Maintenance and service
8.6 Replacement of parts subject to wear and tear
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
68
63 18 351 D3576
D3576.201.01.05.02 11.2018
CAUTION
Make sure that the small metal cylinder of the cable is properly placed
in the anti-pull protection. Fiber in cable may break if not correctly
mounted resulting in high repair costs.
Only use the Sirona Dental Systems battery pack, see "Spare parts"
[→ 23].
8.6
Replacement of parts subject to wear and tear
Check the following parts subject to wear and tear and replace where
applicable:
● Optical fibers (change after 10 treatments/sterilizations)
● Single-use tips (change after every treatment)
● Handpiece sleeve incl. plastic body and silicone switch coat (change
after 400 treatments/sterilizations)
● Bending tool for single-use tips (change after 100 treatments/
sterilizations)
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
For further informations, see chapter "Cleaning, disinfection and
sterilization" [→ 61].
CAUTION
Only use parts from Sirona Dental Systems, see "Spare parts" [→ 23].