2 Safety information
2.8 Laser protective goggles
Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
10
63 18 351 D3576
D3576.201.01.05.02 11.2018
2.8
Laser protective goggles
All persons present in the room (patient, dentist and assistant) must
always wear the laser protective goggles delivered along with the
SIROLaser Xtend.
Check before use
Before using the laser protective goggles, please read and observe the
instructions for use provided by the manufacturer and attached to the
goggles in the case. Make sure that the laser protective goggles:
● are not damaged
● conform to standard EN 207 with protection level LB5
● are suitable for the correct wavelength (labeled on the goggles)
These instructions apply particularly when using goggles supplied from
an outside source that are not included in the scope of delivery of the
SIROLaser Xtend.
Optical instruments
Never use optical instruments such as microscopes, eye loupes or
magnifiers together with the original protective goggles. Otherwise
sufficient eye protection can no longer be ensured.
2.9
Optical fiber and single-use tip
Optical fiber and connection socket
Make sure that no dust, dirt and foreign particles can enter the optical
fiber socket or the optical system. Never place your finger or any other
objects in the optical connectors. Otherwise the unit may be
permanently damaged.
When disconnecting the optical fiber from the SIROLaser Xtend, always
cover the connection socket with the special protection cap supplied.
Make sure that the optical system is clean before connecting the optical
fiber.
The optical fiber must not be twisted. There is a risk of breakage.
Single-use tips
Switch the SIROLaser Xtend off immediately if the optical fiber is
broken. Otherwise the single-use tips may become hot.
The single-use tips must be checked for sure seating prior to each use.