Sirona Dental Systems GmbH
Operating Instructions SIROLaser Xtend
3 System description
3.4 Symbols and abbreviations
63 18 351 D3576
D3576.201.01.05.02 11.2018
17
After upgrade
SIROLaser Xtend
λ
= 970 nm ± 15nm
P
= 10 W
= 635 - 650 nm
P
= 1 mW
IEC 60825-1:2007
max
max
λ
λ
= 970 nm ± 15nm
P
= 14 W
= 635 - 650 nm
P
= 1 mW
IEC 60825-1:2007
max
max
λ
Specification of laser output power and wavelength of IR and aiming
beam, see also chapter "Technical data" [→ 19].
CAUTION
CLASS 4 VISIBLE AND INVISIBLE LASER
RADIATION WHEN OPEN. AVOID EYE
OR SKIN EXPOSURE TO DIRECT
OR SCATTERED RADIATION
Warns of potential laser radiation hazards when opening the laser unit.
CAUTION
VISIBLE AND INVISIBLE LASER RADIATION.
AVOID EYE OR SKIN EXPOSURE TO
DIRECT OR SCATTERED RADIATION
CLASS 4 LASER PRODUCT
Warns of Class 4 laser radiation hazards when using the unit.
Warns of laser radiation emission at the distal tip of the handpiece.
FINGER SWITCH
Warns of laser radiation hazards when the fiber connector is
unscrewed.
Laser Stop
"Laser Stop" key: press this button in case of an emergency
Use only with power supply
Sinpro MPU 100-106
Operate the unit exclusively with the Sinpro MPU100-106 power supply
134°C
The disassembled handpiece sleeves and the fibers may be sterilized
only in autoclaves with saturated water vapor at minimum sterilization
values of 134°C (273.2°F), 3 min. holding time and 2.04 bar (29,59 psi)
overpressure.
Single use only for the bendable tip
Refers to directive 2002/96/EC and EN 50419
Do not dispose with domestic waste