SAFETY STANDARDS AND CLASSIFICATION OF THE SYSTEM
Page 30 of 37
SICAT JMT
⁺
Jaw Motion Analysis System
Benzstrasse 1A
85551 Kirchheim
Detailed information on EMC values and manufacturer data is provided in the tables for this handbook
chapter.
Medical electrical devices are subject to special safety measures regarding EMC (electromagnetic
compatibility) and have to be installed and put into operation pursuant to the instructions provided
below.
CAUTION
Although the SICAT JMT
+
system corresponds with the requirements of the DIN
EN 60601-1-2 norm in terms of all points, mobile phones cannot be excluded
from influencing the jaw motion analysis system. If possible, such devices should
not be operated during measurements in the environment of the JMT
measurement system.
CAUTION
The use of accessories, especially connection cables to the PC that are not
supplied by SICAT for the SICAT JMT
+
system, or are expressly approved for use
with the device, can lead the JMT systems to have increased emissions or
reduced interference immunity.
CAUTION
The SICAT JMT
+
system should not be used near X-ray systems, engines or
transformers with a big connected load, for example, as electrical or magnetic
interference fields can distort the correct measurements and/or make them
impossible. The same applies to neighboring power cables and devices that do
not have CE marking. If operation is required in the immediate proximity of
sources of interference, it is necessary to observe the device under all
circumstances to make sure it is operating in the proper way in this
arrangement.
GUIDELINES AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC EMISSIONS
The measurement systems in the JMT range of products are intended for operation in the
electromagnetic environment stated below. The customer or user of the JMT measurement system
should ensure that it is operated in such an environment.
Emissions tests
Compliance
Electromagnetic environment - guideline
RF emissions according
to CISPR 11
Group 1
The measurement systems in the JMT range of
products only use RF energy for their internal
functioning. For this reason their RF emissions are
very low and it is unlikely that neighboring
electronic devices will suffer interference.
RF emissions according
to CISPR 11
Class B
The JMT measurement system is suitable for use
in all facilities, including those in domestic
settings and those directly connected to the
public power supply network that supplies
buildings that are used for domestic purposes.
Harmonic emissions
according to IEC
61000-3-2
Class B
Voltage fluctuations/
flicker according to
IEC61000-3-3
Compliance