PRODUCT DESCRIPTION
Page 12 of 37
SICAT JMT
⁺
Jaw Motion Analysis System
ultrasonic transmitters. The accompanying head bow consists of six and/or eight ultrasonic
microphones. Both modules are connected with the evaluation electronics in the SICAT JMT
+
measurement system via a connecting cable.
During operation, the ultrasonic transmitters continuously emit impulses, from the operating time of
which, between the transmitter and receiver microphones, the evaluation electronic calculates the
absolute room coordinates of the markers using a triangulation method.
The calculation of the measurement coordinates and further measurement parameters, as well as the
compensation of the disturbance variables, occurs on a PC-supported basis in the evaluation programs.
BLOCK DIAGRAM OF THE MEASUREMENT SYSTEM
CONTROL ELEMENTS AND CONNECTIONS
VIEW FROM FRONT
1
2
Head bow
LJ-sensor
3.4