FUNCTIONAL CHECKS, PREPARATION, DISPOSAL
SICAT JMT
⁺
Jaw Motion Analysis System
Page 25 of 37
CHECK LIST FOR THE RECORDING OF ERROR MESSAGES
NOTICE
To be able to provide you with the optimum support in the case of operational
faults to your SICAT JMT
+
system, our service team employees require the
following information:
■ Serial numbers of the SICAT JMT
+
system and lower jaw sensor/head bow
– The serial numbers are on the name plates on the rear of the measurement system and/or on the
cables for the head bow and LJ-sensor.
■ SICAT JMT
+
software version
■ Operating system version of your measurement PC
– e.g. Windows 7 Professional Service pack 1 (to find under Windows 7: Windows Start button >
Control Panel > System)
■ Further components connected to the measurement system
– Video camera, EMG amplifier cable
■ List of all USB/Bluetooth devices connected to the measurement PC
– e.g. mouse, printer, other measuring systems etc.
■ Screen shot of the error message, or exact wording
– e.g. "Timeout reading from USB"
■ Accurate as possible description of the process that led to the error message
– e.g. "Type A" measurement started, button "B" then clicked, movement "C" then completed,
changed over to function "D", upon changing back, the error message "xyz" was reported, etc.
PREPARATION METHODS
NOTICE
After every case of use of the SICAT JMT
+
system, a re-preparation is required
according to DIN EN ISO 17664. All accessory parts that come into contact with
the patient's mucosa have to be sterilized before use.
CAUTION
Before starting cleaning work or disinfection, switch the measurement system
off under all circumstances and disconnect it completely from the power supply
network.
The following accessory parts are only intended for one-off use on a patient, and should not be prepared
again after use.
■ Occlusal attachment
■ Bite fork type SI
MANUAL CLEANING
■ Before sterilization, clean the accessory parts by hand under running water (drinking water quality,
30 °C ± 5 °C, flow rate 2 liters/min.) with a medium strength toothbrush for 30 seconds.
■ Complete the sterilization immediately subsequent to the cleaning.
■ The cleaning of measurement systems and electrical accessories (head bow, LJ-sensor, foot switch,
IR remote control) should only be carried out when the system is switched off and the charger and/or
USB cable are unplugged, and using a damp cloth.