AREA OF USE AND SAFETY
SICAT JMT
⁺
Jaw Motion Analysis System
Page 7 of 37
SICAT DATA EXPORT
The XML export function enables the use of jaw movements determined with the SICAT JMT
+
system in
CAD/CAM systems and PACS systems for the functional optimization of dentures and of mouth-guards. A
connecting cup serves as an interface for the referencing. It carries reference marks that can be captured
by imaging systems such as surface scanners or DVT.
SAFETY
ENVIRONMENTAL CONDITIONS
The SICAT JMT
+
jaw motion analysis system is suitable for use in dry interior rooms, as can be found in
clinics, medical practices and laboratories.
Permissible operating temperature: 10°C to 40°C
Relative humidity: 30 % to 70 %
CAUTION
The device should not be operated in wet zones, damp areas (swimming baths,
saunas) or climatic chambers.
The measurement systems are not intended for use in areas where there is a risk
of explosion, rooms used for medical purposes or in a flammable atmosphere
(oxygen-enriched).
The devices should not be used near engines or transformers with a big
connected load, for example, or heavy current power lines, as electrical or
magnetic interference fields can distort the correct measurements and/or make
them impossible.
To avoid reciprocal faults from occurring, two SICAT JMT
+
systems should never
be operated in the same room or near other ultrasound emitting devices (e.g.
ultrasonic cleaners, bird scare devices, alarm systems), as this can cause the
measured values to be falsified.
STORAGE AND TRANSPORT
The storage and transport of the measurement system should only be carried out in the original
packaging, as provided by SICAT.
Storage temperature: -20°C to +70°C
Relative humidity: 5% to 90%
Air pressure: 700 hPa to 1060 hPa