- transmitter pulse
- upper layer limit (red line)
- sphere echo
- lower layer limit (red line)
- echo from the additional weight (if used)
- bottom echo
all well separated, and no other echoes within the layer limits.
TS-measurement
Select the TS-Detection Menu to get the horizontal projection window. With the
sphere in the center of the beam, the TS compensated and the TS uncom-
pensated should be identical. These values are read on the screen underneath
the horizontal projection window. If there is a small difference use the TS
compensated value as the measured TS value. It is recommended that the
measured TS value is logged together with other important information. A
recommended form is attached at the end of this document.
If the measured TS value differs from the known TS value of the sphere, then
calculate a new TS transducer gain:
TS measured - TS sphere
New transd. gain = Old transd. gain +
_________________________________________
2
Select the new TS Transducer Gain in the Transceiver Menu and check that the
measured target strength is correct.
In software version 5.20 and higher, TS values have been corrected for TVG
inaccuracies because of the sample intervals and the pulse rise time.
s
A
- measurement
The calibration of the TS-measurement described in the previous paragraph is
the primary calibration and it will in many cases be a sufficient calibration. The
TS-measurement, however, is based on the peak value of the echo samples in
the sphere echo, whereas the s
A
-measurement is based on integration
(averaging) of the echo samples. The received echo may have a smoothed rise
and decay. Therefore the algorithm for calculation of s
A
in the sounder uses an
effective pulse length rather than the nominal pulse length. A test, and if
necessary, a calibration of the s
A
calculation may be carried out according to the
following procedure.
Содержание EY500
Страница 1: ...6LPUDG 3RUWDEOH VFLHQWLILF HFKR VRXQGHU...
Страница 2: ......
Страница 20: ...System familiarization 14 P3400 B...
Страница 44: ...Operation 10 P3402 C Blank page...
Страница 52: ...Operation 18 P3402 C Blank page...
Страница 53: ...Simrad EY 500 P3402 C 19 Figure 4 Scope plot showing 300 samples of transceiver data...
Страница 140: ...Simrad EY 500 20 P3403E A...
Страница 164: ...Maintenance Figure Transceiver with PC...
Страница 165: ...500 senes echo sounders Vlew transceiver...
Страница 166: ...Maintenance Figure interconnection...
Страница 170: ...Maintenance Figure...
Страница 174: ...Maintenance...
Страница 179: ...500 series portable echo sounders Figure tra nsce i ver...
Страница 183: ...500 series echo sounders power...
Страница 185: ...500 series portable echo sounders...
Страница 209: ...Simrad EK 500 EY 500 P2260 E 7 WINCH 2 WINCH 3 WINCH 1 CD481 Figure 1 Rigging of a vessel for sphere calibration...
Страница 224: ...Calibration 22 P2260 E Figure 8...
Страница 225: ...Simrad EK 500 EY 500 P2260 E 23 Figure 9...
Страница 232: ...Calibration 30 P2260 E Blank page...
Страница 234: ...Calibration 32 P2260 E Blank page...
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......