Calibration
14
P2260/E
Split beam
transducer
Single beam
transducer
µ
d
Minimum
distance
r
The sphere after
being moved to
the centre of the
single beam
The sphere in the
centre of the
split beam
(CD3144)
Figure 5
The Scope mode may be useful during this process as well. The scope plot starts
at the beginning of the super layer, and the horizontal axis contains 200 depth
samples. Accordingly, this depth scale is different for the different frequencies.
Check that the sphere is well within the super layer. Select the Test Menu,
Scope and a dynamic range for the plot (50 dB). Then select the highest
amplitude value to be included in the plot (-50 dB). The sphere echo should now
appear on the scope plot. Reduce the dynamic range to give a high vertical
resolution and adjust the highest amplitude to give a convenient plot where the
peak sphere echo reaches the horizontal centre line on the scale. Then move the
sphere in the transversal and longitudinal directions until the maximum
amplitude position has been found.
Use the TS detection window for single beam to read maximum TS value.
Look at the echogram and check that you have, from the top:
Содержание EY500
Страница 1: ...6LPUDG 3RUWDEOH VFLHQWLILF HFKR VRXQGHU...
Страница 2: ......
Страница 20: ...System familiarization 14 P3400 B...
Страница 44: ...Operation 10 P3402 C Blank page...
Страница 52: ...Operation 18 P3402 C Blank page...
Страница 53: ...Simrad EY 500 P3402 C 19 Figure 4 Scope plot showing 300 samples of transceiver data...
Страница 140: ...Simrad EY 500 20 P3403E A...
Страница 164: ...Maintenance Figure Transceiver with PC...
Страница 165: ...500 senes echo sounders Vlew transceiver...
Страница 166: ...Maintenance Figure interconnection...
Страница 170: ...Maintenance Figure...
Страница 174: ...Maintenance...
Страница 179: ...500 series portable echo sounders Figure tra nsce i ver...
Страница 183: ...500 series echo sounders power...
Страница 185: ...500 series portable echo sounders...
Страница 209: ...Simrad EK 500 EY 500 P2260 E 7 WINCH 2 WINCH 3 WINCH 1 CD481 Figure 1 Rigging of a vessel for sphere calibration...
Страница 224: ...Calibration 22 P2260 E Figure 8...
Страница 225: ...Simrad EK 500 EY 500 P2260 E 23 Figure 9...
Страница 232: ...Calibration 30 P2260 E Blank page...
Страница 234: ...Calibration 32 P2260 E Blank page...
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......