Calibration
26
P2260/E
Select the Layer Menu/Layer-1 Menu and set:
Type:
Pelagic
Range (layer thickness L)
See table below
Range Upper:
See table below
Sv Threshold
-100 dB
No. of Sublayers:
1
Frequency
kHz
Range L
m
Range Start
m
2TL
dB
18
50
3975
96
38
20
990
80
120
10
195
61
200
5
147,5
59,5
Range settings and transmission loss for different frequencies
Start the noise - speed trial with the vessel in a fixed position. Make a short
echogram recording and an integration table for the same interval. Print a
marker line when starting the recording and give the line a number that at
the same time is logged in the noise-speed report form at the end of this
document.
Print a second marker line at the end of the recording, connect the propeller
and increase the speed to 2 knots. When the vessel has reached a stable
speed, make a new echogram recording interval and integration table for the
same interval. Continue the same procedure at 2-knot intervals until
reaching maximum speed. Use marker lines to separate the different
intervals. When at maximum speed, disconnect the propeller as quick as
possible and print marker lines for each 2 knots as the speed decreases
towards 0.
If the noise level quickly decreases towards the level at 0 knots as soon as the
propeller has been disconnected, this means that the noise is generated by
the propeller. If the recording at decreasing speed is more or less equal to the
same noise level at increasing speed, the cause is probably flow-noise.
If the increase in level is propeller noise and this substantially reduces the
performance of the sounder, repeat the tests with different combinations of
pitch and propeller speed if possible, to find the most favourable conditions.
When making echogram recordings at different speed levels, select the Test
Menu/Transceiver and read the noise power P
N
re 1 W. Enter readings in the
same form at the end of this appendix.
Содержание EY500
Страница 1: ...6LPUDG 3RUWDEOH VFLHQWLILF HFKR VRXQGHU...
Страница 2: ......
Страница 20: ...System familiarization 14 P3400 B...
Страница 44: ...Operation 10 P3402 C Blank page...
Страница 52: ...Operation 18 P3402 C Blank page...
Страница 53: ...Simrad EY 500 P3402 C 19 Figure 4 Scope plot showing 300 samples of transceiver data...
Страница 140: ...Simrad EY 500 20 P3403E A...
Страница 164: ...Maintenance Figure Transceiver with PC...
Страница 165: ...500 senes echo sounders Vlew transceiver...
Страница 166: ...Maintenance Figure interconnection...
Страница 170: ...Maintenance Figure...
Страница 174: ...Maintenance...
Страница 179: ...500 series portable echo sounders Figure tra nsce i ver...
Страница 183: ...500 series echo sounders power...
Страница 185: ...500 series portable echo sounders...
Страница 209: ...Simrad EK 500 EY 500 P2260 E 7 WINCH 2 WINCH 3 WINCH 1 CD481 Figure 1 Rigging of a vessel for sphere calibration...
Страница 224: ...Calibration 22 P2260 E Figure 8...
Страница 225: ...Simrad EK 500 EY 500 P2260 E 23 Figure 9...
Страница 232: ...Calibration 30 P2260 E Blank page...
Страница 234: ...Calibration 32 P2260 E Blank page...
Страница 239: ......
Страница 240: ......
Страница 241: ......
Страница 242: ......
Страница 243: ......
Страница 244: ......
Страница 245: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 249: ......
Страница 250: ......
Страница 251: ......
Страница 252: ......
Страница 253: ......
Страница 254: ......
Страница 255: ......
Страница 256: ......
Страница 257: ......
Страница 258: ......
Страница 259: ......
Страница 260: ......