background image

Flow rate exceeds display capability

• Open Collector pulse rate exceeds

maximum of 300 pulses per minute.

• Pulse width set too wide.

K-factors cannot be set to 0.

• Hysteresis value too large
• Defective transmitter

Troubleshooting

Display Condition

Suggested Solutions

Suggested Solutions

• Increase Flow units timebase
•  Move flow decimal one place to the right

•  Increase Output Volume setting
•  Decrease Output pulse width
•  Reduce system flow rate

Enter K-factor between 0.0001 to 99999

•  Change the hysteresis value
• Replace transmitter

“-  -  -  -  -”

“Pulse Overrun Output1”
“Pulse Overrun Output2”

“Value must be more than 0”

Open Collector output is always
activated

‡ SIGNET

Signet Scientific Company, 3401 Aerojet Avenue, El Monte, CA  91731-2882  U.S.A.  •  Tel. (626) 571-2770  •  Fax (626) 573-2057
For Worldwide Sales and Service, visit our website:  www.gfsignet.com  •  Or call (in the U.S.):  (800) 854-4090

GEORGE FISCHER ‡ Piping Systems

3-8550.090-3/(F-4/03) English

 © Signet Scientific Company 2002

Printed in U.S.A. on recycled paper

Ordering Information

Mfr. Part No.

Code

Description

3-8550-1

159 000 047

Flow transmitter, Field mount

3-8550-1P

159 000 048

Flow transmitter, Panel mount

3-8550-2

159 000 049

Flow transmitter, Field mount with relays

3-8550-2P

159 000 050

Flow transmitter, Panel mount with relays

3-8550-3

159 000 051

Flow transmitter, Field mount with dual input/output

3-8550-3P

159 000 052

Flow transmitter, Panel mount with dual input/output

Accessories

Mfr. Part No.

Code

Description

3-8050

159 000 184

Universal mounting kit

3-8051

159 000 187

Flow Integral Mnt NPT

3-8050.395

159 000 186

Splashproof rear cover

3-8050.396

159 000 617

RC Filter kit (for relay use)

3-0000.596

159 000 641

Heavy duty wall mount bracket

3-5000.598

198 840 225

Surface Mount Bracket

3-5000.599

198 840 224

5 x 5 inch adapter plate for +GF+ SIGNET retrofit

3-9000.392

159 000 368

Liquid tight connector kit for rear cover (includes 3 connectors)

3-9000.392-1

159 000 839

Liquid tight connector kit, NPT (1 piece)

3-9000.392-2

159 000 841

Liquid tight connector kit, PG13.5 (1 piece)

7300-7524

159 000 687

24 VDC Power Supply 7.5 W, 300mA

7300-1524

159 000 688

24 VDC Power Supply 15 W, 600mA

7300-3024

159 000 689

24 VDC Power Supply 30 W, 1.3 A

7300-5024

159 000 690

24 VDC Power Supply 50 W, 2.1 A

7300-1024

159 000 691

24 VDC Power Supply 100 W, 4.2 A

Содержание 8550-3

Страница 1: ...sor Inputs Range 0 5 to 1500 Hz Sensor power 2 wire 1 5 mA 5 VDC 1 3 or 4 wire 20 mA 5 VDC 1 Optically isolated from current loop Short circuit protected Flow Flow1 6 25 GPM Flow2 9 75 GPM ENTER 3 8550 090 3 WARNING Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Specifications 2 Installation 3 Electrical Connections...

Страница 2: ...325 Ω max 18 V 600 Ω max 24 V Terminals 11 16 Flow sensor inputs Terminals 7 10 Open collector Outputs Two transistor outputs programmable see CALIBRATE menu as High or Low setpoint with adjustable hysteresis Pulse based on Volume Frequency based on flow rate May be disabled Off if not used Terminals 1 and 2 AUXILIARY power Used only if the flow sensor requires more than 1 5 mA current For Signet ...

Страница 3: ...4 VDC PLC or Recorder Channel 3 4 20 mA in Channel 4 4 20 mA in Transmitter 1 Terminals Transmitter 2 Terminals PLC Terminals Channel 1 4 20 mA Power Supply 12 24 VDC Channel 2 4 20 mA Channel 2 4 20 mA Channel 1 4 20 mA Channel 1 4 20 mA in Channel 2 4 20 mA in Power Supply 12 24 VDC NC NC Terminals Snsr 2 Gnd SHIELD Snsr 2 IN RED Snsr 2 V BLACK Snsr 1 Gnd SHIELD Snsr 1 IN RED Snsr 1 V BLACK 16 1...

Страница 4: ...a pulse that is relative to the flow volume or to the flow rate Low Output triggers when the flow rate is less than the setpoint The output will relax when the flow rate moves above the setpoint plus the hysteresis value High Output triggers when the flow rate is greater than the setpoint The output will relax when the flow rate drops below the setpoint plus the hysteresis value Frequency Output i...

Страница 5: ...UP UP UP DOWN keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with UP or DOWN arrow keys Step 4 Press RIGHT ARROW key to select menu item to be edited The first display element will begin flashing Step 5 Press UP or DOWN keys to edit the flashing element RIGHT ARROW key advances the flashing element Step 6 Press ENTER key to...

Страница 6: ...t source for Open Collector output 1 Flow sensor 1 Flow sensor 2 or Delta Flow Select the mode of operation for Open Collector output 1 Options available are High Low Pulse volumetric or Frequency based on rate The signal may be disabled if not in use In Low or High Mode this Open Collector output will be activated when the Flow rate reaches this value Be sure to modify this setting if you change ...

Страница 7: ...ust be entered to reset the resettable totalizer Adjust the minimum and maximum current output The display value represents the precise current output Adjustment limits 3 80 mA 4 00 mA 5 00 mA 19 00 mA 20 00 mA 21 00 mA Use this setting to match the system output to any external device Active HIGH This setting is used to turn a device pump valve ON at the setpoint Active LOW This setting is used t...

Страница 8: ... 000 047 Flow transmitter Field mount 3 8550 1P 159 000 048 Flow transmitter Panel mount 3 8550 2 159 000 049 Flow transmitter Field mount with relays 3 8550 2P 159 000 050 Flow transmitter Panel mount with relays 3 8550 3 159 000 051 Flow transmitter Field mount with dual input output 3 8550 3P 159 000 052 Flow transmitter Panel mount with dual input output Accessories Mfr Part No Code Descriptio...

Отзывы: