8550-3 Flow Transmitter

‡ SIGNET

Calibrate Menu

Display

(Factory settings shown)

Description

The first three characters set the Flow Rate units of measure. They have no effect on
calculations.  They may be any alpha or numeric character, upper or lower case.  The last
character sets the Flow rate Timebase. Select S (seconds, M (minutes, H (hours) or D (days)

This setting tells the transmitter how to convert the input frequency from the flow sensor into a
flow rate. The K-factor is unique to the sensor model and to the pipe size and schedule.
Refer to data in the sensor manual for the correct value.  Limits:  0.0001 to 99999. (The
K-factor cannot be set to 0.)

This setting identifies the Totalizer units. It has no effect on any calculation. It serves as a
label only. Each character can be any alpha or numerical selection, upper or lower case.

This setting tells the transmitter how to convert the flow sensor signal into a volumetric total.
It also is used as the basis for the Open Collector pulse mode.
The setting is usually the same as the Flow K-factor, or different by x10 or x100.
Limits:  0.0001 to 99999. (The K-factor cannot be set to 0)

Select the input source to be associated with Loop output #1:
Flow sensor #1, Flow sensor #2, or Delta Flow.

Select the minimum and maximum values for the Current loop output #1.
The 8550 will allow any values from 0.0000 to 99999.

Select the input source for Open Collector output #1: Flow sensor #1, Flow sensor #2, or
Delta Flow.

Select the mode of operation for Open Collector output #1.
Options available are High, Low, Pulse (volumetric), or Frequency (based on rate).
The signal may be disabled if not in use.

In Low or High Mode, this Open Collector output will be activated when the Flow rate reaches
this value.  Be sure to modify this setting if you change the Flow Units.

The Open Collector output will be deactivated at Setpoint 

±

 Hysteresis, depending on High or

Low Setpoint selection.  (See details on page 4.)

In Pulse mode, Open collector output #1 will generate one pulse when this volume of flow
passes the sensor. (The measurement is based on the Total1 K-factor).
The 8550 will allow any value from 0.0001 to 99999.

In Pulse mode, this setting defines the duration of Open Collector output pulse #1.  The 8550
allows any value from 0.1 seconds to 999.9 seconds.

In Frequency mode, Open Collector output #1 will simulate the sensor frequency, divided
by this setting.  Set for 1 (input frequency = output frequency).  Set for even numbers
(2, 4, 6, 8 . . . 254 maximum) to scale output fequency.

Use this “note pad” to record important dates, such as annual recertification or scheduled
maintenance.

Flow1 Units:

GPM

>

Flow1 K-Factor:

60.000

>

Total1 Units:

Gallons

>

Total1 K-Factor:

60.00

>

Loop1 Source:

Flow1

>

Loop1 Range: GPM

000.00 

 100.00

>

Output1 Source:

Flow1

>

Output1 Mode:

Low

>

Output1 Setpnt:

       1.0      GPM >

Output1 Hys:

       5.0      GPM >

Output1 Volume:

100.00 Gallons

>

Output1 PlsWdth:

0.1 Seconds

>

Output1 Freq.:
Divide by         1 >

Last CAL:

06-30-01

All functions labeled "1" will repeat for channel 2.

Содержание 8550-3

Страница 1: ...sor Inputs Range 0 5 to 1500 Hz Sensor power 2 wire 1 5 mA 5 VDC 1 3 or 4 wire 20 mA 5 VDC 1 Optically isolated from current loop Short circuit protected Flow Flow1 6 25 GPM Flow2 9 75 GPM ENTER 3 8550 090 3 WARNING Remove power to unit before wiring input and output connections Follow instructions carefully to avoid personal injury Contents 1 Specifications 2 Installation 3 Electrical Connections...

Страница 2: ...325 Ω max 18 V 600 Ω max 24 V Terminals 11 16 Flow sensor inputs Terminals 7 10 Open collector Outputs Two transistor outputs programmable see CALIBRATE menu as High or Low setpoint with adjustable hysteresis Pulse based on Volume Frequency based on flow rate May be disabled Off if not used Terminals 1 and 2 AUXILIARY power Used only if the flow sensor requires more than 1 5 mA current For Signet ...

Страница 3: ...4 VDC PLC or Recorder Channel 3 4 20 mA in Channel 4 4 20 mA in Transmitter 1 Terminals Transmitter 2 Terminals PLC Terminals Channel 1 4 20 mA Power Supply 12 24 VDC Channel 2 4 20 mA Channel 2 4 20 mA Channel 1 4 20 mA Channel 1 4 20 mA in Channel 2 4 20 mA in Power Supply 12 24 VDC NC NC Terminals Snsr 2 Gnd SHIELD Snsr 2 IN RED Snsr 2 V BLACK Snsr 1 Gnd SHIELD Snsr 1 IN RED Snsr 1 V BLACK 16 1...

Страница 4: ...a pulse that is relative to the flow volume or to the flow rate Low Output triggers when the flow rate is less than the setpoint The output will relax when the flow rate moves above the setpoint plus the hysteresis value High Output triggers when the flow rate is greater than the setpoint The output will relax when the flow rate drops below the setpoint plus the hysteresis value Frequency Output i...

Страница 5: ...UP UP UP DOWN keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with UP or DOWN arrow keys Step 4 Press RIGHT ARROW key to select menu item to be edited The first display element will begin flashing Step 5 Press UP or DOWN keys to edit the flashing element RIGHT ARROW key advances the flashing element Step 6 Press ENTER key to...

Страница 6: ...t source for Open Collector output 1 Flow sensor 1 Flow sensor 2 or Delta Flow Select the mode of operation for Open Collector output 1 Options available are High Low Pulse volumetric or Frequency based on rate The signal may be disabled if not in use In Low or High Mode this Open Collector output will be activated when the Flow rate reaches this value Be sure to modify this setting if you change ...

Страница 7: ...ust be entered to reset the resettable totalizer Adjust the minimum and maximum current output The display value represents the precise current output Adjustment limits 3 80 mA 4 00 mA 5 00 mA 19 00 mA 20 00 mA 21 00 mA Use this setting to match the system output to any external device Active HIGH This setting is used to turn a device pump valve ON at the setpoint Active LOW This setting is used t...

Страница 8: ... 000 047 Flow transmitter Field mount 3 8550 1P 159 000 048 Flow transmitter Panel mount 3 8550 2 159 000 049 Flow transmitter Field mount with relays 3 8550 2P 159 000 050 Flow transmitter Panel mount with relays 3 8550 3 159 000 051 Flow transmitter Field mount with dual input output 3 8550 3P 159 000 052 Flow transmitter Panel mount with dual input output Accessories Mfr Part No Code Descriptio...

Отзывы: