SC5305A Operating & Programming Manual
Rev 2.1.0
49
A
P P E N D I X
A
-
S
P E C I F I C A T I O N S
Definition of Terms
The following terms are used throughout this datasheet to define specific conditions:
Specification (spec)
Defines guaranteed performance of a calibrated instrument under the
following conditions:
Three hours storage at room temperature (standardized to 25 °C)
followed by thirty minutes minimum warm-up operation.
Specified environmental conditions are met within the specified
operating temperature range of 0 °C to 40 °C unless otherwise noted.
Recommended calibration intervals are used.
Typical data (typ)
When used with <, > or in a range, defines performance met by
approximately 80% of all instruments manufactured. This data is not
guaranteed, does not include measurement uncertainty, and is valid only
at room temperature (standardized to 25 °C).
Nominal values (nom)
Characterizes product performance by means of average performance of
a representative value for the given parameter (e.g. nominal impedance).
This data is not guaranteed and is valid only at room temperature
(standardized to 25 °C).
Measured values (meas)
Characterizes expected product performance by means of measurement
results gained from individual samples.
Specifications are subject to change without notice. For the most recent product specifications, visit
www.signalcore.com
.