EEPROM Data Content
The following list describes the data contents of the EEPROM in detail. All addresses shown are the
starting offset positions in the EEPROM, and are the starting address for block of data.
Manufacturing Information
(0x00).
This is an unsigned integer value that contains information used by
the factory for production purposes.
Product Serial Number
(0x04).
This is an unsigned integer value that contains the SC5305A serial
number. It is unique for every product produced. It is used for the purpose of tracking the history of the
product.
RF Module Serial Number
(0x08).
This is the serial number of the shielded RF metal enclosure containing
the analog and RF circuitry. All calibration data are stored on the EEPROM within the enclosure.
Calibration data are written to this EEPROM at the factory are tracked using the RF module serial
number for the SC5305A.
Product Manufacture Date
(0x0C).
This is an unsigned integer: byte 3 is the Year, byte 2 is the Month,
byte 1 is the day of the month, and byte 0 is the hour of the day.
Last Calibration Date
(0x10).
This is an unsigned integer: byte 3 is the Year, byte 2 is the Month, byte 1
is the day of the month, and byte 0 is the hour of the day.
Firmware Revision
(0x2C).
This is a float 32 value containing the firmware revision.
LO Hardware Revision
(0x30).
This is a float 32 value containing the local oscillator hardware revision.
SC Hardware Revision
(0x34).
This is a float 32 value containing the signal chain hardware revision.
Calibration Temperature
(0x50).
This is a float 32 value containing the temperature at which the device
was calibrated.
TCXO DAC Value
(0x54).
This is an unsigned integer containing the value for the reference DAC to adjust
the precision of the temperature-compensated crystal oscillator (TCXO).
YIG Calibration
(0x64).
Data is reserved for device use.
IF Filter Bandwidths
(0x184, 0x188).
These two float 32 data points contain the filter bandwidths of
IF3_FILTER0 and IF3_FILTER1, respectively. These are only available if the product contains non-standard
filters, different from those provided with the base product.
Gain / Temperature Coefficients
(0x1A0).
This is a 3x8 float matrix, where data is concatenated by rows,
that is, data is read back row by row. These coefficients, derived during calibration, are needed to
compute for gain as a function of temperature. They are 2
nd
order polynomial coefficients and are
measured over eight different frequencies. See “Gain Correction” section for more information on gain
correction factors. Table 6 is an example of the coefficient data and their format. Variables
and
are the first and second order coefficients.