SC5305A Operating & Programming Manual
63
General Specifications
Environmental
Operating temperature
(33)
................................................................................................................ 0 °C to +40 °C
Storage temperature
(34)
................................................................................................................ -40 °C to +70 °C
Operating relative humidity
(35)
...................................................................... 10% to 90%, non-condensing
Storage relative humidity
(35)
............................................................................. 5% to 90%, non-condensing
Operating shock
(36)
................................................................................ 30 g, half-sine pulse, 11 ms duration
Storage shock
(36)
.................................................................................... 50 g, half-sine pulse, 11 ms duration
Operating vibration
(37)
.................................................................................................. 5 Hz to 500 Hz, 0.31 g
rms
Storage vibration
(37)
...................................................................................................... 5 Hz to 500 Hz, 2.46 g
rms
Altitude .............................. 2,000 m maximum (maintaining 25 °C maximum ambient temperature)
Physical
Dimensions (W x H x D, max envelope) .................................................................................. 1.6” x 5.2” x 8.4”
PXIe/cPCIe form factor ............................................................................................................................... 3U, 2-slot
Weight ...................................................................................................................................................................... 2.1 lbs
Power consumption .............................................................................................................................. 12 V @ 2.1 A
3.3 V @ 0.2 A
Communication interface .............................................................................................................. PXIe (PCIe bus)
Safety
............................................................................................................. Designed to meet the requirements of:
IEC 61010-1, EN 61010-1, UL 61010-1, CSA 61010-1
Electromagnetic compatibility (EMC)
.........................................
Designed to meet the requirements of:
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity 1, EN 55011 (CISPR 11)
Group 1, Class A emissions, AS/NZS CISPR 11: Group 1, Class A emissions, FCC 47 CFR Part
15B: Class A emissions, ICES-001: Class A emissions
CE
............................................................................................................................................. Meets the requirements of:
2014/35/EU; Low-Voltage Directive (safety), 2014/30/EU; Electromagnetic Compatibility
Directive (EMC Directive)
Warranty
.................................................... 1 year parts and labor on defects in materials or workmanship
(33) Meets requirements of IEC-60068-2-1 and IEC-60068-2-2. Operating temperature may be
extended to +55 °C with appropriate user-provided cooling solution. Contact SignalCore for
recommended minimum airflow rates.
(34) Meets requirements of IEC-60068-2-1 and IEC-60068-2-2.
(35) Meets requirements of IEC-60068-2-56 and MIL-PRF-28800F, Class 3.
(36) Meets requirements of IEC-60068-2-27 and MIL-PRF-28800F, Class 3.
(37) Meets requirements of IEC-60068-2-64 and MIL-PRF-28800F, Class 3.