4-16KR
Use as a robot-controlled Centrifuge
50
Version 10/2010, Rev. 1.9 of 02/07/2012
Translation of the original operating manual
Commands for centrifuges with motor driven lid or hatch
close[CR]
closes the lid / hatch
Commands for centrifuges with rotor positioning
setpos n[CR]
n=0: unlock the rotor
n>0: go to position n
the lid must be close for positioning
If the centrifuge is running, this command stops the run automatically
and the rotor goes to position n. If the centrifuge is equipped with a
hatch in the lid, the hatch opens automatically during positioning.
pos[CR]
Outputs the position of the rotor in
positioning mode
run n[CR]
Starts the centrifuge with speed n [rpm]
If the centrifuge is equipped with a hatch in the lid, this command
closes the hatch and the centrifuge begins to start the run when the
hatch is closed.
Commands for Rotor Cycle Counter
An additional rotor cycle counter is implemented for free use by the
user. This counter can only be read out by serial interface. The
maximum count value is 4294967295. The value will be set to 0 in
case of overflow.
Commands:
rcycle[CR]
Displays the current rotor cycle counter.
bcycle[CR]
Displays the current bucket cycle counter.
ONLY Spincontrol L
erasercycle[CR]
Resets the rotor cycle counter to “0”.
NOT Spincontrol L