SSG5000A Programming Guide 47
3.5.2.24 Target Level (:SENSe[:POWer]:SPC:TARGet)
Command Format
:SENSe[:POWer]:SPC:TARGet <power>
:SENSe[:POWer]:SPC:TARGet?
Instruction
Set the nominal level expected at the input of the sensor
Get the nominal level expected at the input of the sensor
Parameter Type
Float, unit: dBm, dBuV, uV, mV, V, nW, uW, mW, W, Default: dBm
Parameter Range
-
120 dBm ~ 20 dBm
Return
Float, unit: dBm
Default
0 dBm
Menu
SENSOR > Level Control > Target Level
Example
SENSe:SPC:TARGet -6
SENSe:SPC:TARGet?
3.5.2.25 Level Limit ([:SOURce]:POWer:LIMit)
Command Format
[:SOURce]:POWer:LIMit <power>
[:SOURce]:POWer:LIMit?
Instruction
Set the upper limit for the RF output power
Get the upper limit for the RF output power
Parameter Type
Float, unit: dBm, dBuV, uV, mV, V, nW, uW, mW, W, Default: dBm
Parameter Range
-
120 dBm ~ 20 dBm
Return
Float, unit: dBm
Default
0 dBm
Menu
SENSOR > Level Control > Level Limit
Example
POWer:LIMit 1
POWer:LIMit?