40 SSG5000A Programming Guide
Instruction
Delete the selected row in the flatness list.
Parameter Type
Integer
Parameter Range
Less than the total count of the flatness.
Return
None
Default
None
Menu
LEVEL > Flatness > Delete
Example
CORRection:FLATness:DELete 0
3.5.2.9 Flatness List Count ([:SOURce]:CORRection:FLATness:COUNt?)
Command Format
[:SOURce]:CORRection:FLATness:COUNt?
Instruction
Indicates the total count of the number of elements in the flatness
correction table
Parameter Type
None
Parameter Range
None
Return
Integer
Default
0
Menu
LEVEL > Flatness
Example
CORRection:FLATness:COUNt?
3.5.2.10 Flatness List Store ([:SOURce]:CORRection:STORe)
Command Format
[:SOURce]:CORRection:STORe <file_name>
Instruction
Save the correction data in the list
Parameter Type
String
Parameter Range
None
Return
None
Default
None
Menu
LEVEL > Flatness > Save