SSG5000A Programming Guide 49
3.5.2.28 Catch Range (:SENSe[:POWer]:SPC:CRANge)
Command Format
:SENSe[:POWer]:SPC:CRANge <power>
:SENSe[:POWer]:SPC:CRANge?
Instruction
Set the capture range of the control system
Get the capture range of the control system
Parameter Type
Float, unit: dB
Parameter Range
0 dB ~ 50 dB
Return
Float, unit: dB
Default
0 dB
Menu
SENSOR > Level Control > Catch Range
Example
:SENSe:SPC:CRANge 10
:SENSe:SPC:CRANge?
[:SOURce]:SWEep Subsystem
3.5.3.1 Sweep State ([:SOURce]:SWEep:STATe)
Command Format
[:SOURce]:SWEep:STATe OFF|FREQuency|LEVel|LEV_FREQ
[:SOURce]:SWEep:STATe?
Instruction
Activate frequency or/and level sweep
Parameter Type
Enumeration
Parameter Range
OFF|FREQuency|LEVel|LEV_FREQ
Return
Enumeration
Default
OFF
Menu
SWEEP > Sweep State
Example
:SWEep:STATe OFF
:SWEep:STATe?