SSG5000A Programming Guide 39
POWer:ALC?
3.5.2.6 Flatness List State ([:SOURce]:CORRection[:FLATness])
Command Format
[:SOURce]:CORRection[:FLATness] ON|OFF|1|0
[:SOURce]:CORRection[:FLATness]?
Instruction
Activate/deactivate flatness correction list.
Parameter Type
Boolean
Parameter Range
ON|OFF|1|0
Return
Boolean
Default
0
Menu
LEVEL > Flatness
Example
CORRection:FLATness ON
:CORRection?
3.5.2.7 Flatness List Add Row ([:SOURce]:CORRection:FLATness:PAIR)
Command Format
[:SOURce]:CORRection:FLATness:PAIR <freq>,<power>
Instruction
Insert a new row in the flatness list.
Parameter Type
Float, Float
Parameter Range
Freq: Full freq range
Power: Full power range
Return
None
Default
None
Menu
LEVEL > Flatness > Add
Example
CORRection:FLATness:PAIR 3 MHz,-3
3.5.2.8 Flatness List Delete Row ([:SOURce]:CORRection:FLATness:DELete)
Command Format
[:SOURce]:CORRection:FLATness:DELete <row>