SIG
L
ENT
SDS2000X-E User Manual
187
Reference Position
The reference position setting determines the physicial point that the oscilloscope uses
during vertical and horizontal scale changes. In some situations, it is more convenient to
use a fixed position on the display.
Press the
Utility
button on the front panel and then press the
Reference Pos.
softkey to
enter the Reference POS menu
.
Press the
Vertical
softkey to select
Fixed Offset
or
Fixed Position
.
Fixed Position
: When the vertical gain is changed, the oscilloscope will keep the vertical
offset level indicator stationary.
Fixed Offset
: When the vertical gain is changed, the oscilloscope will have the vertical
offset level indicator move with the actual voltage level.
Press the
Horizontal
softkey to select
Fixed Delay
or
Fixed Position
.
Fixed Position
: When the time base is changed, the oscilloscope will keep the horizontal
offset indicator stationary.
Fixed Delay
: When the time base is changed, the oscilloscope will have the horizontal
delay indicator move with the trigger point.
Power On Line
Press the
Utility
button on the front panel; then; press the
Power On Line
softkey to
select
Enable
or
Disable.
Enable:
When the oscilloscope is energized, the oscilloscope will start automatically
without user intervention. This is ideal for situations where physical access to the
instrument power switch is difficult or impossible, such as automated test equipment
or remote locations.
Disable
: This setting requires user intervention to energize the scope. Press the
power key at the lower-left corner of the front panel to start the oscilloscope.
Содержание SDS2000X-E Series
Страница 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Страница 14: ......
Страница 29: ...SIGLENT SDS2000X E User Manual 3 Appearance and Dimensions Figure 1 Front View Figure 2 Top View...
Страница 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Страница 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Страница 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Страница 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Страница 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Страница 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Страница 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Страница 162: ...SIGLENT 136 SDS2000X E User Manual...
Страница 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Страница 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Страница 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...