Update Firmware and Configuration
The firmware and configuration can be updated directly via USB flash drive.
Do the following steps to update the firmware:
1. Insert the USB flash drive which contains the correct firmware and the configure files
into the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Firmware
softkey to open the file manager.
7. Turn the
Universal Knob
to select the update file which should be with an ADS postfix;
and then press the
Press to Update
softkey to start update the firmware. The process
needs about 7 minutes. And during the update, do not remove power to the
oscilloscope, otherwise the oscilloscope may be permanently damaged and may not
restart again.
8. After finish the update, the screen will pop-out the message “
Firmware
decompressed. Please restart and wait…
”.
9. Restart the oscilloscope to finish the firmware update.
Do the following steps to update the configuration:
1. Insert the USB flash drive which contains the correct firmware and the configure files
into the USB host interface on the front panel of the oscilloscope.
2. Press the
Utility
button on the front panel to enter the UTILITY function menu.
3. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
4. Press the
Next Page
softkey to go to the third page of the UTILITY function menu.
5. Press the
Update
softkey to enter the UPDATE function menu.
6. Press the
Configure
softkey to open the file manager.
7. Turn the
Universal Knob
to select the update file ( *.CFG file extension) and then
press the
Press to Update
softkey to start update the firmware. The process needs
about 30 seconds.
8. After finish the update, the screen will pop-out the message “
Firmware
decompressed. Please restart and wait…
”.
9. Restart the oscilloscope to finish the configuration update.
Содержание SDS2000X-E Series
Страница 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Страница 14: ......
Страница 29: ...SIGLENT SDS2000X E User Manual 3 Appearance and Dimensions Figure 1 Front View Figure 2 Top View...
Страница 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Страница 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Страница 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Страница 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Страница 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Страница 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Страница 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Страница 162: ...SIGLENT 136 SDS2000X E User Manual...
Страница 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Страница 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Страница 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...