LIN Trigger
This part will provide a brief introduction and description for the operation of the LIN
trigger.
Trigger conditions:
Break
— The oscilloscope will be triggered at the position of the break field’s
break delimiter.
ID
(Frame ID) — The oscilloscope will be triggered at the position of the identifier
field’s stop bit, if the value of a frame’s ID is equal to specified value.
(
Note:
If the data’s value is 0xXX, any data value will be matched.)
ID + DATA
(Frame ID and Data) — The oscilloscope triggers when a frame with
an ID and data equal to the selected values is detected. Use the
Universal Knob
to select the value for the ID, Data1 and Data2.
1) The ID’s value is the same as set value.
2) If you have set either Data1’s or Data2’s value, and the signal has a data is
the same as that value. If you have set both Data1’s and Data2’s value, the
signal should has two consecutive data, the first data’s value is Data1,
second data value is Data2.
(
Note:
If the data’s value is 0xXX, any data value will be matched.)
Data Error
— The oscilloscope will be triggered when errors (such as ID check
error, checksum error, sync byte field error) are detected.
Operation steps:
1. Press
Setup
to enter the
TRIGGER
function menu.
2. Press
Type
and select
Serial
.
3. Press
Protocol
and select
LIN
.
4. Press
Trigger Setting
softkey to enter
LIN TRIG SET
menu.
5. Press
Condition
and select the trigger condition by
Universal Knob
:
If you
select
ID
condition
:
a. Press
ID
softkey and set its value by turning the
Universal Knob
.
If you
select
ID+DATA
condition
:
a. Press
ID
softkey and set its value by turning the
Universal Knob
.
b. Press
DATA1
softkey and set its value by turning the
Universal Knob
.
c. Press
DATA2
softkey and set its value by turning the
Universal Knob
.
Содержание SDS2000X-E Series
Страница 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Страница 14: ......
Страница 29: ...SIGLENT SDS2000X E User Manual 3 Appearance and Dimensions Figure 1 Front View Figure 2 Top View...
Страница 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Страница 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Страница 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Страница 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Страница 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Страница 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Страница 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Страница 162: ...SIGLENT 136 SDS2000X E User Manual...
Страница 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Страница 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Страница 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...