SDS2000X-E
User Manual
Hamming
These reduce leakage and improve amplitude accuracy. However,
frequency resolution is also reduced.
Flat Top
This window provides excellent amplitude accuracy with moderate
reduction of leakage, but with reduced frequency resolution.
Blackman
It reduces the leakage to a minimum, but with reduced frequency
resolution.
5. Press the
Center
softkey, and then turn the
Universal Knob
to select the desired
center frequency.
6. Press the
Hz/div
softkey, and then turn the
Universal Knob
to select the desired
resolution frequency.
7. Press the
Scale
softkey, and then turn the
Universal Knob
to select the desired
vertical FFT scale.
8. Press the
Ref Level
softkey, and then turn the
Universal Knob
to select the desired
vertical FFT offset.
9. Press the
Unit/Load
softkey to enter Unit/Load menu. Press
Unit
softkey to select the
unit of vertical axis. The units of the vertical axis can be dBVrms, dBm or Vrms which
use a logarithmic scale or a linear scale to display vertical amplitude respectively. If
you need to display the FFT frequency spectrum in a relatively larger dynamic range,
dBVrms is recommended. Press
Ext Load
softkey, and then turn the
Universal Knob
to select the external load value.
10. Press the
Mode
softkey to select
Split
,
Full Screen
or
Exclusive
display mode.
Split:
The source channel and the FFT operation results are displayed separately.
The time domain and frequency domain signals are displayed clearly.
Full Screen:
The source channel and the FFT operation results are displayed in
the same window to view the frequency spectrum more clearly and to perform
more precise measurements.
Exclusive:
The displayed waveforms of all channels are disabled, only the FFT
operation results are displayed in the window. This isolated view allows you to
view the frequency spectrum more clearly and to perform more precise
measurements.
Содержание SDS2000X-E Series
Страница 1: ...User Manual SDS2000X E Series Digital Oscilloscope UM0102E E01A SIGLENT TECHNOLOGIES CO LTD...
Страница 14: ......
Страница 29: ...SIGLENT SDS2000X E User Manual 3 Appearance and Dimensions Figure 1 Front View Figure 2 Top View...
Страница 69: ...SIGLENT SDS2000X E User Manual 43 Figure 13 x Interpolation Figure 14 Sin x x Interpolation...
Страница 98: ...SIGLENT 72 SDS2000X E User Manual Figure 29 Relative Window Trigger...
Страница 100: ...SIGLENT 74 SDS2000X E User Manual Figure 30 Interval Trigger...
Страница 103: ...SIGLENT SDS2000X E User Manual 77 Figure 32 State Dropout Trigger...
Страница 105: ...SIGLENT SDS2000X E User Manual 79 Figure 33 Runt Trigger...
Страница 125: ...SIGLENT SDS2000X E User Manual 99 Figure 49 CAN Trigger...
Страница 129: ...SIGLENT SDS2000X E User Manual 103 Figure 51 LIN Trigger...
Страница 162: ...SIGLENT 136 SDS2000X E User Manual...
Страница 200: ...SIGLENT 174 SDS2000X E User Manual Figure 95 Built in Arb interface...
Страница 206: ...SIGLENT 180 SDS2000X E User Manual Figure 100 WiFi setting menu...
Страница 215: ...SIGLENT SDS2000X E User Manual 189 Figure 108 Option Information...