Siemens
7482255 Rev 01
Page 17 of 52
October 2002
Chapter 4 - Removing and Replacing Spare Parts
4.
Hold the bumper, and remove the front cover from the G50 or the G60 S.
5.
Pry off six screw caps from the top cover, then remove six M4 X 10 flat head
screws and washers as shown in Figure 4-14: Do not remove the two sets
of screws located in the middle.
6.
Pull out the top cover from the G50 or the G60 S as shown in Figure 4-14.
7.
Disconnect all of the cables from the rear side of P91W as shown in
Figure 4-11.
8.
Unscrew the two M3 X 8 binding head screws and external lock washers
from bracket F, then remove the P91W and its mounting plate from the
OEM bay.
9.
Unscrew the four M4 X 8 binding head screws and external lock washers
that attach the OEM bay to the G50 or the G60 S.
10.
Remove the OEM bay from the system.
11.
Detach the plastic bag from the OEM bay, and remove the four M3 X 8 flat
head screws provided in this bag.
12.
Remove the two S-type brackets and the separator from the OEM bay by
unscrewing the four M3 X 8 flat head screws that attach them.
13.
Turn the CP900 upside down.
14.
With the bottom of the CP900 facing upward, place the CP900 on a padded
or protected platform of 25 mm or higher.
15.
Invert the OEM bay and place it over the bottom surface of the CP900.
16.
Using the four M3 X 8 flat head screws removed from the plastic bag,
secure the OEM bay to the CP900 by installing the four screws through the
holes marked with the character C.
17.
Return the inverted CP900 and the attached OEM bay to their original
upright position.
18.
Place the OEM bay, with the CP900 attached, in its original position in the
G50 or the G60 S, but do not locate the OEM bay all the way toward the
rear; confirm that enough space remains between the CP900 and the inter-
face panel to re-establish the cable connections between them.
19.
Reinstall the mounting hardware that was initially removed from the OEM
bay.
20.
Re-install the P91W, attached to its mounting plate, in the OEM bay, and re-
connect the cables to the P91W.
21.
At the P91W, position the number 1 dip switch in the On position.
22.
Fasten the front edge of the mounting plate to the bracket F, using two M3
X 8 binding head screws with external lock washers that were removed.
23.
At the CP900, set the RGB IMPEDANCE switch to the 75 _/HIGH position
and the SYNCHRONIZATION SIGNAL IMPEDANCE switch to the HIGH
position.
Содержание SONOLINE G50
Страница 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Страница 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Страница 182: ...Page 42 of 42 7482255 Rev 01 Siemens October 2002 Service Screens and Menus Figure 8 32 Color Scale Pattern ...
Страница 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Страница 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Страница 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Страница 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Страница 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Страница 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Страница 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Страница 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...