
Page 12 of 22
7482255 Rev 01
Siemens
October 2002
Rebooting the System Software
Observe the screen during the reboot to detect HDD error messages. If
errors are detected, run the Scan Disk application from the
Start
→
Pro-
grams
→
Accessories
→
System Tools
menu. If no errors are detected
on the HDD, reboot the system.
4.
If DIMAQ-IP still does not function properly, load new DIMAQ-IP software.
4.2 Corrupted Images and Clips
If, when REVIEW is selected, corrupted images or clips are displayed, the
following failures can be the cause.
1.
New images or clips produced by real-time scanning are corrupted, but
images loaded from studies on the HDD or from a CD are not corrupted.
Corruption of ultrasound images is caused by a failure in the host module
or the failure of the image transfer path (serial). Inspect and test the host
module and the serial path; replace the host module if necessary.
2.
New images from real-time scanning are not corrupted, but images
recalled from the HDD or from a CD are corrupted.
The corruption of images loaded from the HDD or a CD is caused by the
DIMAQ-IP software or by the CD-R/W drive software. Reload the DIMAQ-
IP software to resolve the problem.
5.0 Rebooting the System Software
Use the following instructions to reboot the system software after a system
failure.
Powering down and re-energizing the G50 or the G60 S will perform a
complete (cold) boot of all subsystem software.
The following instructions describe the re-boot of the system software only (soft
or warm boot). Use this procedure if the system locks up.
1.
Press and hold down the New Patient key.
2.
Locate the key bearing the small square on its key cap: This key is located
above the VCR key.
3.
Press and hold down the key bearing the small square on its key cap.
4.
Press and hold down the Freeze key: Press and hold the
New Patient
+
Square
+
Freeze
keys to perform a soft boot of the system software.
Note:
The re-boot of the DIMAQ-IP is described in "Review Switch Failure"
Содержание SONOLINE G50
Страница 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Страница 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Страница 182: ...Page 42 of 42 7482255 Rev 01 Siemens October 2002 Service Screens and Menus Figure 8 32 Color Scale Pattern ...
Страница 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Страница 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Страница 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Страница 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Страница 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Страница 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Страница 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Страница 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...