Page 4 of 8
7482255 Rev 01
Siemens
October 2002
Installing the Version 1.2 Software Update
23.
Verify that the stored files appear on the diskette: Service Data Backup,
MMDDYYGenSysPre, MMDDYYOBTabFor, MMDDYYDefSet, and MMD-
DYYTotSysCon.
24.
Select Exit.
25.
Eject and store the MO diskette.
1.2 Installing the Version 1.2 DIMAQ-IP Software Update
Use the following steps to install the version 1.2 DIMAQ-IP software update.
1.
Energize the G60 S.
2.
Press the
Review
key to display the Study screen.
3.
Press the
F11
key.
4.
At the keyboard, enter
eng
(three letters in lower case).
5.
Press the
F11
key to display a DIMAQ-IP version window.
6.
Record the DIMAQ-IP version information.
7.
Select the Cancel button to exit the DIMAQ-IP version window.
8.
If the DIMAQ-IP version is 1.2, the version 1.2 DIMAQ-IP update need not
be installed: Continue with "Installing the Version 1.2 System Software
Update" (page 7-6).
9.
Back up all exam and patient data to a CD before continuing with the
DIMAQ-IP software update.
10.
Open the upper-rear panel of the system to expose the Microsoft Windows
label attached to the DIMAQ-IP.
11.
Record the 20 characters of the Microsoft Windows 2000 Product Identifi-
cation code; this license code is required to enable the new software.
12.
Close the upper-rear panel.
13.
Press the
F4
key and select the
Service
menu item.
14.
Enter the service password and select the Ok button to display the Main
Menu screen.
!
Caution:
The original release of the G60 S contained version 1.0 system soft-
ware and version 1.022 DIMAQ-IP software. The Speed Info 011/2002/US
describes the update of the version 1.022 DIMAQ-IP software to version 1.023.
On existing G60 S systems, the version 1.0 system software will operate accept-
ably with both the version 1.023 DIMAQ-IP software and the version 1.2
DIMAQ-IP software; to avoid problems
, update the version 1.022 or 1.023
DIMAQ-IP software to version 1.2,
before
updating the system software to ver-
sion 1.2.
!
Caution:
Update of the DIMAQ-IP software will over-write all data on the HDD.
Before beginning the installation of the DIMAQ-IP software update, back up all
customer exam and patient data to a CD.
Содержание SONOLINE G50
Страница 4: ...Page iv 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 10: ...Page x 7482255 Rev 01 Siemens October 2002 G50 and G60 S Service Manual ...
Страница 22: ...Page 12 of 12 7482255 Rev 01 Siemens October 2002 Mechanical and Electrical Specifications ...
Страница 140: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Verifying the Operation of the Software ...
Страница 182: ...Page 42 of 42 7482255 Rev 01 Siemens October 2002 Service Screens and Menus Figure 8 32 Color Scale Pattern ...
Страница 190: ...Page 8 of 8 7482255 Rev 01 Siemens October 2002 Concluding the Maintenance ...
Страница 196: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 Figure B 1 Installation Report Form ...
Страница 214: ...Page 18 of 18 7482255 Rev 01 Siemens October 2002 G60 S Spare Parts Photographs ...
Страница 216: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...
Страница 222: ...Page 6 of 6 7482255 Rev 01 Siemens October 2002 ...
Страница 226: ...Page 4 of 4 7482255 Rev 01 Siemens October 2002 ...
Страница 228: ...Page 2 of 6 7482255 Rev 01 Siemens October 2002 This page intentionally left blank ...
Страница 234: ...Page 2 of 2 7482255 Rev 01 Siemens October 2002 ...