Requirement
● The terminal area of the cable socket is suitable for cables with diameters ranging from 6 to
12 mm.
● These cables use stranded wires with 1 mm
2
as single conductors ("+", "-" and ground).
● You are using a crimping tool from HARTING (article number 09 99 000 0110).
Procedure
1. Slide the sleeve and the screwed joint on the cable.
2. Strip approx. 8 mm of the cable ends.
3. Crimp the contact parts on the cable ends.
4. Assemble the cable socket.
8
+
,
$
Connector pin assignment with Han 7D or
Han 8D plug or cable socket
I
A
Output current
U
H
Auxiliary power
6.4
Connect M12 cable socket to the cable
WARNING
Loss of safety required for approval by using the M12 device plug.
You may only use the plug for non-Ex devices; otherwise, the safety required for the approval
is not guaranteed.
Note
A conductive connection must not exist between the shield and the connector housing.
Note
Observe the protection class of the M12 device plug when defining the protection class.
In devices where a plug is already mounted on the enclosure, the connection is made via a
cable socket.
1. Thread the parts of the cable socket as described by the manufacturer of the cable socket.
2. Strip approximately 18 mm of the bus cable
①
.
Connecting
6.4 Connect M12 cable socket to the cable
SITRANS P320/P420 with 4 to 20 mA/HART
72
Operating Instructions, 11/2019, A5E44852162-AC
Содержание SITRANS P
Страница 10: ...Table of contents SITRANS P320 P420 with 4 to 20 mA HART 10 Operating Instructions 11 2019 A5E44852162 AC ...
Страница 272: ...Abbreviations SITRANS P320 P420 with 4 to 20 mA HART 272 Operating Instructions 11 2019 A5E44852162 AC ...
Страница 280: ...Index SITRANS P320 P420 with 4 to 20 mA HART 280 Operating Instructions 11 2019 A5E44852162 AC ...