Technical data
11.7 Pressure - temperature ratings
FC410 with Modbus
96
Operating Instructions, 02/2016, A5E33120874-AC
Image 11-1 Metric flange ratings, EN 1092-1 (P: Process pressure; T: Process temperature)
Image 11-2 ANSI flange ratings, ASME B16.5 (P: Process pressure; T: Process temperature)
11.7.1
Stainless steel sensors
Table 11- 13 EN1092-1 [bar]
PN (bar)
Temperature TS (°C)
-50
0
50
100
150
200
16
16.0
16.0
16.0
15.2
13.8
12.7
40
40.0
40.0
40.0
37.9
34.5
31.8
63
63.0
63.0
63.0
59.7
54.3
50.1
100
100.0
100.0
100.0
94.8
86.2
79.5
160
100.0
100.0
100.0
100.0
100.0
100.0
Table 11- 14 ISO228-G and ASME B1.20.1 NPT [bar]
PN (bar)
Temperature TS (°C)
-50
0
50
100
150
200
100
100.0
100.0
100.0
100.0
100.0
100.0
Содержание SITRANS FC410
Страница 2: ......
Страница 70: ...Commissioning 6 14 Process variables FC410 with Modbus 68 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 92: ...Troubleshooting FAQs 10 2 Troubleshooting FC410 with Modbus 90 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 142: ...Sensor dimension dependent default settings FC410 with Modbus 140 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 146: ...Zero point adjustment FC410 with Modbus 144 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 150: ...CRC calculation FC410 with Modbus 148 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 152: ...Exception codes E 1 Exception handling FC410 with Modbus 150 Operating Instructions 02 2016 A5E33120874 AC ...
Страница 154: ...Float definition F 1 Float definition FC410 with Modbus 152 Operating Instructions 02 2016 A5E33120874 AC ...